cLEANING AND DISINFEcTING

Thoroughly clean all components to remove medication residuals and possible impurities after each treatment.
The compressor (1) and the air hose (6) should be cleaned with a clean, moist cloth.
Always wash your hands well before cleaning and disinfecting the accessories.
Do not expose the compressor to water or heat.
Replace the air hose for each treatment with a new patient or in case of impurities.
Unplug the device before cleaning.
cleaning with water
Wash all nebuliser components (except air hose) under warm tap water (max. 60 °c) for about 5 mi-
nutes adding if necessary a small quantity of detergent following dosage and use limitations as provided by
detergent manufacturer.
Rinse thoroughly making sure that all deposits are washed away and leave to dry.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage
and use limitations as provided by disinfectant manufacturer.
Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.)
or 134 °C (7 min.).
After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.