Herunterladen Inhalt Inhalt Diese Seite drucken

Fault Finding - Stuart SCT1 Bedienungsanweisung

Vorschau ausblenden Andere Handbücher für SCT1:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

right hand corner of the display of the SCT1
temperature controller. The amber control LED
on the front panel of the hotplate will NOT be
illuminated in monitor mode.
NB: When in monitor mode, the SCT1
temperature controller operates as a digital
thermometer only, and no longer controls
the temperature of the medium in use. The
hotplate temperature can now be manually
adjusted to the desired setting.
To change from monitor mode to control
mode
To change from monitor mode to controller
mode, press either the + or – key on the SCT1
temperature controller once. The unit will revert
to controller mode, and will be indicated by the
illuminated amber control LED visible on the
front panel of the hotplate. The red LED on the
top right hand corner of the SCT1 temperature
controller display will extinguish when the unit is
in controller mode.
N.B. For safety reasons, when reverting
from monitor to controller mode the SCT1
temperature controller will default to the
set point temperature of 20°C.

6. Fault Finding

In the event of malfunction the SCT1
temperature controller can self diagnose certain
fault conditions. In this case the display will show
a message in the form of 'Err XX' where XX is
either a single or double digit code, which
identifies the fault.
Error code
Err 1
Err 2
Err 4
Err 8
Err 16
4
Fault condition
Thermometer not calibrated
Electronic fault
Probe plug removed in
control or monitor mode
fault –return to factory
Probe out of sample when
in control mode
Ambient temperature inside
hotplate too high
(controller attached)
Maintenance, Servicing
& Repair
WARNING: Ensure the hotplate is
disconnected from the mains electricity
supply before attempting maintenance or
servicing.
This range of equipment does not require routine
servicing. The only maintenance required is to
clean external surfaces with a damp cloth and
mild detergent. Spillages or splashes should be
cleaned up straightaway after isolating the
hotplate from the mains electricity supply.
6.1 Repairs
Any repairs or replacement of parts MUST be
undertaken by suitably qualified personnel.
Only spare parts supplied or specified by
Bibby Scientific Ltd. or its agent should be
used. Fitting of non-approved parts may
affect the performance of the safety
features designed into the instrument.
For a comprehensive list of parts required by
service engineers conducting internal repairs, or
to receive a service manual, please contact the
Sales Department of Bibby Scientific Ltd quoting
both the model and serial number.
If in any doubt please contact the Technical
Department of Bibby Scientific Ltd.
Remedy
Return to factory
Return to factory
Reconnect probe. Continuing
error code indicates hotplate
Replace probe in sample
Switch off at 'mains' and allow to cool

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis