Operating Instruction
Measuring Temperature
To avoid possible damage to the meter or
other equipment, remember that while the
meter is rated for −20°C to +1000°C and −4°F
to 1832°F , the K Type Thermocouple provided
with the meter is rated to 250°C. For temper-
atures out of that range, use a higher rated
thermocouple.
1. Connect the negative "–" plug of the K type
thermocouple to the COM jack, and the
positive "+" plug to the µAmA°C°F jack.
2. Set the range switch to "°C" or "°F" posi-
tion.
3. Move the end of the thermocouple into
the area to be measured.
4. Wait until the reading is stable, read the
reading on the display.
33
Measuring Capacitance
1. Connect the black test lead to the COM
jack, the red test lead to the
(Note: The polarity of the red lead is posi-
tive "+".)
2. Set the range switch to
3. Connect the test leads across the capaci-
tor to be measured.
4. Wait until the reading is stable, then read
the reading. (For high-capacitance meas-
urements, it may take about 30 seconds
for the reading to be stable.)
Note:
Before measurement, make sure that the
capacitor to be measured has been dis-
charged thoroughly. To improve the accuracy
of low capacitance measurement, subtract
the residual capacitance of the meter and
test leads by using relative measurement.
For the measurements ≥100 µF , the display
will show "OL" as overrange indication.
95980775
EN
jack.
position.