DIODE TEST
1. Insert the RED test lead into the VΩ jack
test lead into the COM jack
switch
to the Continuity/Resistance/Diode test
2
Black lead Red lead
NOTE: The meter will default to continuity testing in this mode.
Press the 'SEL/DC ZERO' button
mode. The Diode icon
2. Touch the test leads to the diode. A reading of 200-800 mV on
the display indicates forward bias. 'OL' indicates reverse bias.
An open device will show 'OL' in both polarities. A shorted
device will show approximately 0 mV.
FREQUENCY/DUTY CYCLE
1. Insert the RED test lead into the VΩ jack
test lead into the COM jack
switch
to the Frequency/Duty cycle
2
Black lead Red lead
NOTE: The meter defaults to frequency testing. To test the duty
cycle, press the 'SEL/DC ZERO' button
appropriate icon (either Hz or %) appears on the display.
1. Measure by connecting the test leads across the circuit.
OPERATING INSTRUCTIONS
4
will appear on the display.
4
5
, and rotate the Function selector
twice to enter Diode testing
9
5
, and rotate the Function selector
once. Ensure that the
9
and the BLACK
setting.
and the BLACK
setting.