17. TROUBLESHOOTING

If your TENS machine is not working properly, please check the following:
Problem
Possible causes
No display
Flat batteries.
Batteries inserted
incorrectly.
Damaged springs in
battery compartment.
Low battery
Low batteries.
display
No sensation Incorrect connection.
Intensity is not strong
enough.
Output will
Open circuit cut-out
not increase
operating.
above zero
Lead not connected to
body or faulty/damaged.
(Lead wires can break at the bend where they leave the machine giving
no, or intermittent, output).
Sudden
If you disconnect and re-
change in
connect a few minutes
sensation
later, the signal will feel
quite a lot stronger.
If the above review has failed to resolve your problem, or to report unexpected
operation or events, call TensCare or your local supplier or distributor (address on
back cover) for advice.
Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are trained to
assist you with most issues you may have experienced, without the need to send your
product in for service.
Solution
Replace batteries.
Remove plastic wrap.
Check + / - .
Contact supplier.
Replace batteries.
Check that you have applied the 4
electrode pads to ensure a complete
circuit.
Increase strength.
Most users will feel something at a setting
below 10.
Ensure machine is attached to your body
correctly.
Try using the second lead wire.
Purchase replacement if necessary.
Always return strength to zero after
disconnecting the leads or electrode pads.