2 Installation

2.1

Technical specifications

Hoffmann Munich-classification-NO: GARANT 380865_1
Dimensions of the device:
Dimensions of the sensor:
Measuring range, resolution and accuracy
"DC X 1" indicate the standard resolution measuring mode and "DC X 10" indicate the high
resolution measuring mode.
*DC X 1
DC X 10
*AC
* The measuring range in the DC X 1 and AC modes are automatically switched over.
** This equipment displays values over the measuring range, but the accuracy warranty applies to the
measuring range only.
(Accuracy)
If the measurement is 123.5mT in the DC X 1 mode (Measuring range 0
0.1mT)±(123.5 X 0.05+3 X 0.1)mT=±6.475mT= ±6.5mT
Power supply:
Operating temperature:
Storage temperature:
Scope of supply:
Measuring magnetic field
21
140 mm x 64 mm x 33 mm
65 mm x 6.0 mm x 1.0 mm
**Measuring range
0~200.0 mT (0~2000G)
200.1~3000.0 mT (2001~30000G)
0~300.00 mT (0~3000.0G)
0~150.00 mT (0~1500.0G)
150.1~300.0 mT (1501~3000G)
301.0~1500.0 mT (3010~15000G)
4 pieces of manganese batteries; 1.5V; R6P
(for 100 hours of continuous operation)
0 °C to 40 °C
–10 °C to 60 °C
Measuring device, sensor, 4 pieces of batteries,
carrying case, instruction manual
DC magnetic field
AC magnetic field
www.hoff mann-group.com
Display
Accuracy
resolution
(20 °C ± 1 °C)
0.1 mT (1G)
± (5% of rdg. + 3digit)
1 mT (10G)
± (5% of rdg. + 10digit)
0.01 mT (0.1G)
± (3% of rdg. + 5digit)
0.01 mT (0.1G)
0.1 mT (1G)
± (5% of rdg. + 20digit)
1 mT (10G)
~
200.0 mT Display resolution
Average detection system
Application frequency 40–500Hz
(Sine wave)
DE
EN
FR
IT
ES
CS
PL
RU
ZH
21