PROBLEM
Device not filling
Unable to
disconnect
portable from
Reservoir
Device frosted
& no pulse or
continuous flow
Device is making a
hissing noise,
(venting)
NOTE:
VENTING IS
NOT A LEAK!
• For oxygen dependent patients (24 hours a day treatment) a
secondary oxygen supply should be available all the time during
therapy.
• If primary oxygen supply fails to supply adequate oxygen, secondary
oxygen supply should be employed for proper oxygen delivery.
PROBABLE CAUSE
1. Not pulling vent-to-
Fill lever.
2. Not connecting fill
connectors
completely.
3. Reservoir empty.
1. Fill connectors frozen
from moisture on fill
connectors.
1. Device overfilled.
1. Device was just filled.
2. Device was turned on
its side.
3. Patient is not using
device after filled.
WARNING
15
REMEDY
1. Pull out vent-to-Fill
lever.
2. Make sure fill
connectors are fully
engaged.
3. Contact liquid Oxygen
Supplier to refill
Reservoir.
1. a) Depress the release
button
on
Reservoir, if equipped.
b) Allow time for device
to warm.
NOTE: If during this
time
liquid
begins to leak from
device, keep away and
contact
Equipment Provider.
1. Wait approximately 30
minutes until Device
returns to normal
operating conditions.
1. Start using Device.
2. Hissing should reduce
in a few seconds.
3. Device will continue to
vent until depleted.
(This is a normal
operation)
the
oxygen
your