• This device is not intended for use by patients who would suffer
immediate, permanent, or serious health consequences as a result of
an interruption in the oxygen supply.
• NEVER smoke in an area where oxygen is being administered.
• NEVER use near any type of flame or flammable/explosive
substances, vapors or atmosphere.
• DO NOT use oils, greases, lubricants or any combustible materials
on or near this product. Wash hands properly prior to usage.
• DO NOT touch liquid oxygen or parts that have been in contact with
liquid oxygen. Liquid oxygen is extremely cold (-297°F/-183°C).
When touched, liquid oxygen, or parts of the equipment that have
been carrying liquid oxygen, can freeze skin and body tissue.
TO AVOID INCREASED RISK OF FIRE
Keep this equipment away from electrical appliances. Use and
store Reservoir and Portable units at least 5 feet (1.52 m) from
electrical appliances that may cause heat or sparks.
Keep oxygen equipment away from open flames. Use and store
Reservoir and Portable at least 5 feet (1.52 m) away from
equipment such as furnaces, water heaters, and stoves that may
contain open flames.
Keep equipment in a well-ventilated area at all times.
These devices periodically release small amounts of oxygen gas
that must be ventilated to prevent buildup. DO NOT store liquid
oxygen equipment in a closet, car trunk, or other confined area.
DO NOT place blankets draperies or other fabrics over
equipment.
High concentrations of oxygen can cause rapid burning of other
substances.
ALWAYS confirm prescribed dose before administering to
patient and monitor on a frequent basis.
• DO NOT carry the Portable device under your clothing. These
devices normally vent oxygen. Wearing a Portable device under
clothing may saturate fabrics with oxygen and cause them to
burn rapidly if exposed to sparks or flame. It may take several
hours for oxygen levels in fabric to return to normal.
ALWAYS keep tubing or oxygen supply line away from path of
walking to avoid potential trip or fall.
DO NOT use if dirt or contaminants are present on or around fill
connectors on the Portable device or Reservoir.
NO OXYGEN is delivered when the Pulse/Flow Selector is at the
"0" Setting.
NO OXYGEN is delivered in between settings.
Set Pulse/Flow Selector to "0" when not in use.
WARNING