4. When your Portable System is set to Pulse Setting P2 thru P6
(EasyMate 6+6 Model), oxygen will be delivered only while you inhale. You
should notice a small pulse at the beginning of each breath. Oxygen is
delivered during this pulse. it is normal for the device to occasionally skip
a breath or to pulse twice in one breath.
5. Breathe through the nose and feel a pulse of oxygen as you inhale.
5a. When your Portable is set to a continuous setting, a constant flow of
oxygen will be delivered.
6. Insert Portable System into the carry bag.
Carry bag may become saturated with oxygen which could cause it to
burn rapidly if exposed to sparks or flames. It may take several hours
for the oxygen levels in fabric to return to normal.
Positioning Device
Always keep the Portable System in an upright
position while in use, as illustrated in (Figure 6).
Placing the device on its front, back or upside down
will shorten the usage time of device.
If the Portable is placed on its front, back or upside down (see below)
for more than one (1) minute, check the liquid oxygen level in the
Portable. By placing the Portable in the positions, listed below may
cause a loss of oxygen, or the vent to fill lever to become too cold for
filling. Set the Portable in an upright position for at least 15 minutes
before attempting to fill.
Front
WARNING
Back
12
Figure 6
Upside
down