To reduce the effect of electromagnetic sources of interference, please read the following
warnings:
L
Do not use any portable television or radio devices in the immediate vicinity of the
nursing bed.
L
Do not use any walkie talkies or mobile phones in the immediate vicinity of the
nursing bed.
L
Should any unintended movements occur, remove the power plug for the nursing bed
from the wall socket.
Guidelines and manufacturer's declaration - Electromagnetic transmissions
The nursing bed is intended for operations in surroundings described below. The customer or bed user
needs to make sure that it will be operated in surroundings of this type.
Electromagnetic
Agreement
radiation
measurements
HF - Radiation
pursuant to CISPR 11
Group 1
(partial)
HF - Radiation
pursuant to CISPR 11
Class B
(partial)
Transmission of
harmonics in accordance
Class A
with
IEC 61000-3-2
Transmission of
Voltage fluctuations/
In accordance with
flickers pursuant to
IEC 61000-3-3
Guidelines and manufacturer's declaration - Electromagnetic resistance
The nursing bed is intended for operations in surroundings described below. The customer or the
user of the nursing bed needs to make sure that it will be operated in surroundings of this type.
Resistance
tests
Discharge of static
± 6 kV contact discharge
electricity (ESD)
pursuant to
± 8 kV Air discharge
IEC 61000-4-2
Quick stray
± 2 kV for power cables
electrical
factors / bursts pursuant
± 1 kV for input and
to
output cables
IEC 61000-4-4
± 1 kV Series mode
Surges
voltage
pursuant to
± 2 kV Common-mode
IEC 61000-4-5
voltage
< 5 % V
(>95 % drop in V
for ½ period
40 % V
Voltage drops,
(60 % drop in V
Short-term disruptions
for 5 periods
And Fluctuations in the
supply voltage
70 % V
pursuant to
(30 % drop in V
IEC 61000-4-11
for 25 periods
< 5 % V
(>95 % drop in V
for 5 seconds
Electromagnetic surroundings - Guidelines
The nursing bed uses HF energy exclusively for its internal
function. Therefore its HF radiation is very low and it is
unlikely that neighboring electronic devices will be
disrupted.
The nursing bed is suitable for use in all facilities which are solely used as
living areas and which have an immediate connection to the
public power supply, which also supplies the building
which is being used as living quarters.
IEC 60601
Compliance level
Test level
± 6 kV contact discharge
± 8 kV Air discharge
± 2 kV for power cables
± 1 kV for input and
output cables
± 1 kV Series mode voltage
± 2 kV Common-mode voltage
< 5 % V
T
T
)
(>95 % drop in V
T
for ½ period
40 % V
T
T
)
(60 % drop in V
T
for 5 periods
70 % V
T
T
)
(30 % drop in V
T
for 25 periods
< 5 % V
T
T
)
(>95 % drop in V
T
for 5 seconds
Page 17 
Luna 2, Luna Basic 2, Luna UL 2
Electromagnetic surroundings -
Guidelines
Flooring needs to be made of wood or
concrete or covered with ceramic tiles
If the floor is covered with synthetic
material, then relative humidity needs to
be at least 30%.
The quality of the power voltage
needs to meet standards typical in a
commercial or hospital environment.
The quality of the power voltage
needs to meet standards typical in a
commercial or hospital environment.
)
T
The quality of the power voltage
needs to meet standards typical in a
commercial or hospital environment.
)
T
Should the user of the nursing bed
require the continuous function even with
the occurrence of disruptions to the
)
T
power supply, it is recommended that the
nursing bed receive undisrupted power
from an emergency source or a battery.
)