Fault
Problem descrip-
number
tion on the display
301
Leak in system
302
System clogged
305
Battery empty
306
Canister full
313
Filter clogged
401
Battery low
402
USB connection
not permitted
405
Standby mode
XXX
Internal fault
20
EN
Troubleshooting on the
display
Check system for leaks
• Check that tubing is clear/
not twisted
• Check canister filling level
Charge battery
Change canister
Change canister and switch
Liberty on
Charge battery
Unplug USB cable
Switch on or off
• Switch Invia Liberty off
and on
• Inform Medela Customer
Service
Operation halts
Remarks/potential cause
of fault
• Tubing not fixed
• Missing or defective seal
between the canister and
Invia Liberty
• Dressing not sealed
properly
• Tubing is kinked or clogged
• Filter in canister contami-
nated
Pressure is maintained if
Invia Liberty is in upright
position
Filter in the canister is
clogged
Remaining time of battery
approx. 30 min.
USB connection during
operation or in patient
environment not permitted
In standby mode after
5 minutes
If fault message repeats, re-
port fault number to Medela
Customer Service
Operation continues
Pressure