Herunterladen Inhalt Inhalt Diese Seite drucken

HEINE iC1 Bedienungsanleitung Seite 8

Vorschau ausblenden Andere Handbücher für iC1:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Setting up smartphone
Insert the smartphone into the appropriate mounting casing so it is firmly
seated. Make sure that the camera is in the correct position.
To avoid any possible malfunctions, please deactivate all services (flight mode).
Images for documentation purposes can be taken with the standard camera
app on your smartphone. Also, please deactivate the internal flash of the
phone.
Operation
The iC1 allows for a rapid change from polarized to non-polarized examination
mode over a toggle switch. The difference is that no immersion fluid is required
in the polarized examination mode. The default setting is the polarized mode.
Turn the dermatoscope on by using the on/off switch (6). Gently place the
instrument so that the lesion is in the center of the contact plate.
For non polarized mode
Prepare the skin by moistening with HEINE dermatoscopy-oil (use a cotton
swab) or disinfectant spray.
Always use the device in combination with the filter inserts. Only use
contact plates from HEINE.
Switch off the device after each examination.
Brightness control
The electronic voltage regulation in the dermatoscope guarantees a constant
brightness.
An electrical conducting connection between smartphone, PC and a
mains power source is not permissible during operation.
Charging
Charge indicator
First insert the power supply into the wall socket or PC and then the micro
USB into the charging port (4).
When the iC1 is charged via USB, the LED will flash green.
When the battery is fully charged, the LED will be constant green.
When the battery reaches a critical charge state, the LED will light up orange.
The device must now be recharged.
Charge the device outside of the patient environment (1.5 m).
Hygienic reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national
standards, laws and guidelines.
Classification according to KRINKO: non-critical
Spaulding Classification USA: non-critical
In the event of suspected contamination, carry out hygienic preparation
of the instrument.
The described cleaning and disinfection measures do not replace the
specific rules applicable for the establishment.
HEINE Optotechnik only approves the resources and procedures named
in these instructions for use.
Cleaning and disinfection may only be carried out by personnel with
sufficient hygienic knowledge.
Observe the instructions of the manufacturer of the reprocessing media.
Do not use spray or immersion disinfection, dripping wet or heavily
foaming tissues.
Do not use ultrasonic reprocessing and don't sterilize.
Do not use reprocessing media including alcohol.
The contact plates have to be cleaned and/or disinfected after each use.
8/48
med 112655
2015-12-10

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis