Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The Veroval
DS 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
®
such an environment.
IMMUNITY test
IEC 60601 test level
3 V
0.15 MHz to 80 MHz
6V in ISM and amateur radiobands between
Conducted
0.15 MHz and 80 MHZ 80% AM at 1kHz
RF IEC
61000-4-6
10 V/m 80 MHz to 2.7 GHz
380 MHz, 27 V/m
Radiated RF
450 MHz, 28 V/m
IEC 61000-4-3
710 MHz, 745 MHz, 780 MHz 9V/m
RF Wireless
810 MHz,870 MHZ, 930 MHz 28 V/m
Communication
Equipment
1720 MHz,1845 MHZ,1970 MHz 28V/m
IEC 61000-4-3
2450 MHz, 28 V/m
5240 MHz,5500 MHz, 5785 MHz 9V/m
Compliance level
N/A
10 V/m
380 MHz, 27 V/m
450 MHz, 28 V/m
710 MHz, 745 MHz, 780 MHz 9V/m
810 MHz,870 MHZ, 930 MHz 28 V/m
1720 MHz,1845 MHZ,1970 MHz 28V/m
2450 MHz, 28 V/m
5240 MHz,5500 MHz, 5785 MHz 9V/m
84
DS 22 should assure that it is used in
®
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the device, inclu-
ding cables, than the recommended separation distance
calculated from the equation applicable to the frequency
of the transmitter.
Recommended seperation distance
3,5
d =
[
]
80 MHz to 800 MHz
E
1
7
d =
[
]
80o MHz to 2,7 Ghz
E
1
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment mar-
ked with the following symbol:
should
a

Werbung