Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The Veroval
DS 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
®
such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst IEC
61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
IEC 60601 test level
Compliance level
± 8 kV
± 8 kV
contact
contact
±2 kV,±4 kV,
±2 kV,±4 kV,
±8 kV,
±8 kV,
±15 kV air
±15 kV air
± 2 kV
for power supply lines
N/A
± 1 kV
for input/output lines
± 1 kV
differential mode
N/A
± 2 kV
common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycle
N/A
70% UT (30% dip in UT) for
25 cycle
<5% UT (>95% dip in UT)
for 5 sec
30 A/m
30 A/m
Electromagnetic environment – guidance
Floor should be wood, concrete or ceramic tile. If floors are covered with
shythetic material, the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels charactertic of a typical
location in a typical commercial or hospital environment.