LEL or PPM
##
HM
HM
rSt
---
Abbildung 13: IR5500 Menübaum Blinkeingang
Flash Input Menu
HM
HM
SE
in
rSd
in: Flash Signal Input Menu
The flash input data is used to diagnose poor
alignment, path interference, aging Source
###
lamp, etc.
rSd: relative standard deviation of reference
flash input intensity.
rEF
### after rSd = 100% x ref stdev / ref average
ref stdev and ref average: standard deviation
T
and average of reference flash intensity.
###
rSd > 60 for ind time sets fault F12
rSd ≤ 60 resets fault F12
M
rEF: Reference flash
### after rEF: average reference flash intensity
rSd and rEF are updated every 30 seconds