Instruction Manual
Digital multimeter 0590.0382
Digital display
Data hold / Reset
Switch for direct current/alternating current,
continuity
Measuring range/Reading switch
Input socket "10 A"
Input socket "COM"
Input socket "VΩmA"
You can carry out the following using
testo 382:
- Measures DC and AC voltage up to
1000 V
- Measures DC and AC current
- DC current measurement up to 10 A
- Resistance measurement up to 30 MΩ
- Continuity test
According to the conformity certificate, the
instruments meet 89/336/EEC guidelines.
Supplied:
The instruments were tested for EMC in the
1 x testo 382 with protection case
frequency range from 150 kHz - 1MHz. The speci-
1 x battery 9V IEC 6LR61
fied parameters cannot be guaranteed in the fre-
2 x measuring leads with test prods (rd/bl)
quency range from 60 MHz-190 MHz.
1 x Instruction manual
Symbols on the instrument or in the
Please read the Instruction manual carefully
Instruction manual:
prior to using the instrument and adhere to
the instructions given.
Caution! Warning about danger point.
If instructions are not adhered to or war-
Refer to Instruction manual.
nings are ignored, serious or life-threatening
injuries to the user or damage to the instru-
Caution! Dangerous voltage
ment may be the consequence.
If working with voltages greater than 120 V
Direct current
(60 V) DC or 50 V (25 V) eff AC precautio-
nary measures must be adhered to in order
Alternating current
to avoid an electric shock. These values are
the limit for contact voltages in accordance
Ground. Do not exceed max. voltage
with DIN VDE (values in brackets apply to
range between input socket and ground.
medical or agricultural sectors).
Safety measures
Prior to every measurement, please ensure
that the measuring leads and the measuring
The instrument and test prods should only
instrument are in perfect working order.
be held by the handles provided for this
The measuring instrument should only be
purpose.
used in the measuring ranges specified.
Avoid all contact with the test prods.
Never measure current in circuits with more
To change a battery or fuse, the multimeter
than 250V DC.
should only be opened by a person who is
familiar with the dangers of electric current.
The measuring instrument should only be
used in mains systems which are fused to
Before opening, ensure that the measuring
max. 16A.
leads are separated from the object being
Please adhere to current accident prevention
measured and that the instrument is swit-
guidelines.
ched off.
Direct sunlight on the instrument should be
This instrument meets the safety require-
avoided. A perfectly functioning instrument
ments in accordance with IEC-1010-1
and a long life is guaranteed if the above
(EN61010-1) Protection Class II and
points are adhered to.
Overvoltage category, Cat III / 600 V for
electric measuring instruments.
Contamination level 2.
These instruments should not be used in
high-energy circuits. It is suitable for mea-
surements in systems of the Overvoltage
Category III in accordance with
IEC 664 (max. 600 V AC/DC, 10 A).
Measuring
testo 382
The measuring range selection switch must
be at the required function before the test
prods are connected to the object being
measured.
Remove the measuring leads from the
object being tested before switching to a
new function or a different measuring range.
The instrument should only be used in a dry
and clean atmosphere. Dirt and moisture
reduce insulation resistances and could lead
to electric shocks in the case of high volta-
ges.
The instrument should only be used in the
measuring ranges specified.
Check if the instrument is working properly
before use (e.g. on a known current or vol-
tage source, see also DIN VDE 0105, Part
1).
Ensure that the measuring leads are in per-
fect order.
The use of makeshift fuses, in particular the
short-circuiting of fuse-carriers is not per-
mitted and may cause the measuring instru-
ment to be destroyed and cause serious
physical injury.
Auto Off - Function
The instrument switches off automatically
after 15 min if not used. A buzzer sounds.
Switch on the instrument again via the "Data
hold / Reset" button.
Voltage measurement
Maximum 1000V AC/DC should be
applied to the input sockets. The
instrument could be damaged and
the user is endangered if these limit values
are exceeded.
Remove the test prods from the object
being measured before selecting a
different measuring range.
1. Set measuring range selection switch to
direct or alternating current (depending
on the type of voltage being measured).
2. Plug the red measuring lead in the
"VΩmA" socket and the black measuring
lead in the "COM" socket.
3. Connect the test prods to the object
being measured. The reading is shown
on the digital display.
Current measurement
Never measure current in circuits
where there are voltages of more
than 250V. Only currents with max.
300mA can be measured in the "VΩmA"
socket and max. 10A in the "10A" socket.
Measuring direct current:
Set the measuring range selection switch to
"µA" or "mA". If the currents are unknown,
always start the measurement in the highest
measuring range and switch to the lower
respective measuring range to achieve the
highest resolution.
Maintenance
Measuring alternating current:
Press
changeover switch. AC appe-
ars on the display. Switch measuring range
The instrument does not require special
selection switch to "µA" or "mA". If the cur-
maintenance if operated in accordance with
rents are unknown, always start the measu-
the instruction manual . All work on the
rement in the highest measuring range and
instrument, apart from changing batteries or
switch to the lower respective measuring
fuses, should only be carried out by authori-
range to achieve the highest resolution.
sed personnel.
Measuring direct and alternating current:
Cleaning
1. Plug red measuring lead in the "10 A" or
If the instrument becomes dirty through
"VΩmA" input sockets and the black
daily use, it can be cleaned with a damp
measuring lead in the "COM" socket.
cloth and a mild domestic cleaning agent.
2. Connect multimeter to item being tested
On no account should aggressive cleaning
and then switch on item being tested.
agents or solvents be used for cleaning pur-
3. Take reading from digital display.
poses.
Regular calibration
The instrument should be calibrated by our
service department so as to maintain the
accuracy of the measured result. We recom-
Resistance measurement
mend annual calibration (Part No.
0520.2310 Multimeter Calibraiton
The test object should be disconnec-
Certificate).
ted from supply. If in doubt, check
by measuring.
Changing the battery
If the battery symbol
top right corner of the display, the battery
1. Set measurement type selection switch to
should be replaced.
resistance measuring range "Ω".
2. Plug red measuring lead in the "VΩmA"
1. Disconnect the instrument from the cir-
socket and the black measuring lead in
cuit.
2. Switch off the instrument.
the "COM" socket.
3. Open the battery compartment at the
3. Connect test prods to the item being
back by unscrewing the screws.
tested.
4. The resistance value can be read on the
4. Remove the empty battery.
digital display.
5. Insert the new battery, type 9V IEC
6LR61, observe polarity and close the
Continuity test
housing.
The test object should be disconnec-
ted from supply. If in doubt, check
Measurements can be carried out.
by measuring.
Please think of the environment. Do not
throw used batteries in domestic garbage.
Please dispose of responsibly.
1. Set measurement type selection switch to
resistance measuring range "Ω".
2. Press
changeover switch until the
symbol appears in the display.
3. The resistance value can be read from the
The battery should be removed if is
digital display. A buzzer sounds when
resistance values are < 20 Ω.
batteries leak inside the instrument,
return the instrument to us for cleaning
and checking purposes.
Diode test
The test object should be disconnec-
Changing fuses
ted from supply. If in doubt, check
If a fuse has been triggered by overloading
by measuring.
or incorrect handling, change the fuse.
1. Set measuring range selection switch to
Only use fuses with the specified cur-
diode test
2. Press changeover button
until
symbol appears in the display.
3. Plug the red measuring lead in the
1. Disconnect instrument from circuit.
"VΩmA" socket and the black measuring
2. Switch off the instrument and open the
lead in the "COM" socket.
battery compartment as described in
4. Connect test prods to object being tested.
"Changing the battery" and remove the
5. The drop in conducting-state voltage can
battery.
be read on the digital display.
3. Open the instrument by unscrewing the
screws at the back of the instrument.
Note:
4. Remove defect fuse and replace with a
The polarity of the test voltage at the
new fuse, same type.
"VΩmA" socket is "+".
5. Put testo 382 back together again.
Fuse values:
Fuse 1: 250 V / 0.5 A quick-acting
Fuse 2: 250 V / 10 A slow
The use of makeshift fuses, in particular the
short-circuiting of fuse-carriers is not per-
mitted and may cause the instrument to be
destroyed and cause serious physical injury.
Technical data
Display: ..................3-1/2 digit LCD
Analog display:........Bar display with 33
Display range: ........3200 digits
Meas. range selection: Automatic
Battery status
display:....................Battery symbol
Overvoltage
category: ................CAT III / 600 V
Contamination
level:........................2
Power supply: ........9 V IEC 6LR61
Power consumption: Approx. 2.6 mA (typical)
Dimensions (lxwxh): 160 x 80 x 40
+ + - -
appears in the
Weight:....................286 g (with prot. case)
Operating temp.: ....0...+40 °C (0...80 %RH)
Storage temp.:
Range
Direct current voltage (DCV)
300mV
3V
30V
300V
750V
1000V
Alternating current voltage (ACV)
3V
30V
300V
750V
1000V
R R e e s s i i s s t t a a n n c c e e ( ( Ω Ω )
300Ω
not planned to use the instrument
3kΩ
for a longer period of time. If the
30kΩ
300kΩ
3MΩ
30MΩ
Continuity
Range
Signal
Diode
Resolution
1mV
rent and voltage values, switch-off
properties and dimensions.
D D i i r r e e c c t t c c u u r r r r e e n n t t ( ( D D C C A A ) )
300µA
3mA
30mA
300mA
10A
A A l l t t e e r r n n a a t t i i n n g g c c u u r r r r e e n n t t ( ( A A C C A A ) )
300µA
3mA
30mA
300mA
10A
segments
appears (<7 V)
-20...+60 °C (0...80%RH)
Resolution
Accuracy
100uV
±1.2%rdg+ 2dgts
1mV
±0.5%rdg+ 2dgts
10mV
±1.2%rdg+ 2dgts
100mV
±1.2%rdg+ 2dgts
1V
±1.5%rdg+ 2dgts
1V
±1.5%rdg+ 2dgts
50Hz-60Hz
1mV
±2.0%rdg+ 3dgts
10mV
±2.0%rdg+ 3dgts
100mV
±2.0%rdg+ 3dgts
1V
±2.0%rdg+ 3dgts
1V
±2.0%rdg+ 3dgts
Up to 500 Ω
0.1Ω
±1.2%rdg+ 3dgts
1Ω
±1.2%rdg+ 3dgts
10Ω
±1.2%rdg+ 3dgts
100Ω
±1.2%rdg+ 3dgts
1KΩ
±1.2%rdg+ 3dgts
10KΩ
±3.0%rdg+ 5dgts
Up to 500 Ω
Frequency
Signal from
4.1 kHz
<20W ± 10W
Up to 500 Ω
Max.
Accuracy
0.8mA
±8.0%rdg + 2dgts
0.1µA
±1.5%rdg+ 2dgts
1µA
±1.5%rdg+ 2dgts
10µA
±1.5%rdg+ 2dgts
100µA
±1.5%rdg+ 2dgts
10mA
±2.5%rdg+ 2dgts
0.1µA
±2.0%rdg+ 4dgts
1µA
±2.0%rdg+ 4dgts
10µA
±2.0%rdg+ 4dgts
100µA
±2.0%rdg+ 4dgts
10mA
±3.0%rdg+ 4dgts