Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

10.2 EMV-Daten

EMV-Allgemeine Daten
Emission
RF-Emission following
EN 55011
RF-Emission following
EN 55011
Generation of mains harmonics
following IEC 61000-3-2
Generation of mains harmonics
following IEC 61000-3-3
EMV-Tabelle 202
Susceptibility
ESD IEC 61000-4-2
Bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage drops etc
IEC 61000-4-11
H-field at 50/60Hz
IEC 61000-4-8
EMV-Tabelle 204
Susceptibility
Conducted rf
IEC 61000-4-6
Radiated rf
IEC 61000-4-3
EMV-Tabelle 206
Output
power
transmitter
W
0,01
0,1
1
10
100
Technische Änderungen vorbehalten
IEC 60601-test level
±6kV cd
±8kV ad
±2kV mains
±1kV I/O
±1kV dm
±2kV cm
Reduction to
5% for 10ms/ positive
amplitude
5% for 10ms/ negative
amplitude
40% for 100ms
30% for 500ms
0% for 5000ms
3A/m
IEC 60601-test level
3Veff
150kHz to 80 MHz
3Veff
80 MHz to 2,5 GHz
of
Safety distance depending on frequency/m
150kHz to 80MHz
0,12m
0,37m
1,17m
3,69m
11,67m
80MHz to 800MHz
0,12m
0,37m
1,17m
3,69m
11,67m
- 32 -
Übereinstimmung
group 1
Class B
Class A
none
actual level
±6kV cd
±8kV ad
±2kV mains
±1kV I/O
±1kV dm
±2kV cm
Reduction to
5% for 10ms/ positive
amplitude
5% for 10ms/ negative
amplitude
40% for 100ms
30% for 500ms
0% for 5000ms
3A/m
Actual level
3V
3V/m
800MHz to 2,5 GHz
0,24m
0,74m
2,34m
7,38m
23,34m