ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacture's declaration - electromagnetic emissions
The AD-2021 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the AD-2021 should assure that it is used in such an
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/
fl icker emissions IEC
61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The AD-2021 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the AD-2021 should assure that it is used in such an
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic fi eld
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
56
For all ME EQUIPMENT and ME SYSTEMS
environment.
Compliance
Group 1
Class B
Not
applicable
Not
applicable
For all ME EQUIPMENT and ME SYSTEMS
environment.
IEC 60601 test
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic
environment-guidance
The AD-2021 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electro-
nic equipment
The AD-2021 is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings
used for domestic purposes.
Compliance
level
environment-guidance
Floors should be wood,
concrete or ceramic tile. If
fl oors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Power frequency magnetic
fi elds should be at levels
characteristic of a typical loca-
tion in a typical commercial or
hospital environment.
Electromagnetic