The device needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the
requirements of CISPR.
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of
this device.
Other devices that have RF transmitter or source may affect this device (e.g. cell
phones, PDAs, and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Emission tests
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage
Fluctuations
Emissions IEC 61000-3-3
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Health
Monitor should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Compliance
Group 1
Class B
Class A
/
Flicker
Complies
IEC60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV line(s) to
line(s)
Electromagnetic environment - guidance
The device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV line(s) to
line(s)
32
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.