4.14. Parameter calibration

* The instrument is calibrated before shipment, but some specifications may change a bit lot during long
time of use. To ensure the accuracy, the instrument should be calibrated termly.
* Users may regain the accuracy of the instrument by operating the keyboard to calibrate the main
specifications without removing the cover of the instrument.
4.14.1. Enable calibration:
* After booting, the calibration is in off state, and the generator could not be calibrated without inputting
calibration password, this is a way to protect calibrated parameters which may be changed carelessly.
* To enable calibration, select sine wave and then press【Shift】【Cal】-key, the calibration password
displayed as 0, input calibration password 1900, press 【N】-key to enable calibration.
4.14.2. Parameters calibration:
* Press 【Menu】-key to display calibration value on the left, and calibration sequence number on the right
when setting calibration conditions automatically.
* Adjust calibration value to calibrate present selected calibration option and make the output expected.
* Continue to press 【Menu】-key and the calibration sequence number will increase step by step, users
could calibrate all those options respectively, which is shown in the following list.
* During calibration process, press 【Cal】-key at any time then press 【Menu】-key to return the
calibration sequence number to 00.
Parameter calibration table
Sequence
Default calibration
No.
00
01
02
03
04
05 ~ **
** P 4055: Sequence-no.: 05~07
P 4060: Sequence-no.: 05~24
4.14.3. Disable calibration:
* After finishing the calibration, press 【Cal】 key and there display 1900, press any numeric keys then
【N】 key to store the calibration parameters, disable calibration and exit the process.
* During the calibration process, if wrong calibration occurred, press 【Freq】 key at any time to disable
calibration and exit without storing calibration parameters.
* After rebooting, the generator automatically recalls and uses the calibration parameters stored during last
calibration.
Output nominal
value
2047
870
873
300
0,71 Vrms
500
100
Adjust the calibration value till the output
value
0 VDC
output DC voltage as: -20 ~ 20 mVDC
10 VDC
output DC voltage as: 9.88 ~ 10.12 VDC
7 Vrms
output AC voltage as: 6.928 ~ 7.072 Vrms
output AC voltage as: 0,701 ~ 0,719 Vrms
1 MHz
output frequency as: 1 MHz ± 20 Hz
5 Vpp
output amplitude as. 4,5 Vpp ~ 5,5 Vpp
-51-
is within the error range
Zero calibration:
Offset calibration:
Amplitude calibration:
Amplitude calibration:
Frequency calibration:
Flatness calibration: