Delete Data; Calibration; Autozero - Compur Monitors Dositox Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3.3

Delete Data

If you do not want to download the data in the memory of the Dositox, they must be deleted before
starting a new measurement cycle.
? Switch the instrument „OFF".
? Press button „Fct".
? The display will now show „del Data".
? Press the button „Cal" and hold.
? The display will now show a countdown from 5 to 0 before erasing all data in the memory.
4.

Calibration

4.1

Autozero

? The automatic zero must only be executed in clean air. This function is only active if no
data are in the memory of the Dositox.
? Switch the instrument in position „Test".
? Press key „Cal".
? The display will show flashing „000" and then „GCal".
? Now switch the instrument „ON" or „OFF".
? The autozero is finished.
? If the sensor is not yet ready for autozero Dositox will display Err7. Wait some time and
repeat autozero.
27