11 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
The displayed weight does not glow.
The displayed weight is permanently
changing
-
The weighing value is obviously
wrong
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
24
Possible cause
• The balance is not switched on.
• The batteries are empty
• Draught/air movement
• Sites with vibration.
• Loading fork is in contact with foreign
matter.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• Load fork has not been pumped to
reference height
• The control pin is not in position
„Reference height".