"SENSE" Function (HPS-11560 Only)
The HPS-11560 has an automatic voltage control for the rear high-current output. For this, two separate
measuring cables are connected parallel to the connecting cables. The potential drop which occurs on
the connecting cables is measured on these two measuring cables. The laboratory power unit automati-
cally compensates for this voltage drop so that the actually set voltage is supplied to the consumer.
Proceed as follows for connection:
Always connect the supply cables from the power unit to the consumer first. Pay attention to correct
polarity.
Press the clamp release on the rear SENSE connection inwards with a small screwdriver and insert the
cables into the terminal openings. Check that they are plugged firmly.
Now connect the two "SENSE" cables to the consumer observing the correct polarity. The conductor
cross-section for the "SENSE" cables must be at least 0.34mm2.
Always slacken the connection in the reverse order (first of all "SENSE" cables and then the connecting
cables).
Make sure that you contact the SENSE cables as close as possible to the
connecting point of the consumer. Observe correct polarity.
Never short-circuit the "SENSE" cables.