Herunterladen Inhalt Inhalt Diese Seite drucken

Nissei WSK-1011 Handbuch Seite 15

Vorschau ausblenden Andere Handbücher für WSK-1011:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 16
RF emissions CISPR 11
Class B
Note: Operate by DC Power
Harmonic emissions IEC 61000-3-2
N/A
Voltage fluctuations/flicker
N/A
emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity -
WSK-1011 is intended for use in the electromagnetic environment specified below. The customer or the user of
WSK-1011 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
±6 kV contact
(ESD) IEC 61000-4-2
±8 kV air
Electrical fast transient/
±2 kV for power supply lines
burst IEC 61000-4-4
±1 kV for input/output lines
Surge IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
Voltage dips, short
<5% U
(>95% dip in U
) for
T
T
interruptions and
0,5 cycle
voltage variations on
40% U
(60% dip in U
) for 5
T
T
power supply input
cycles
lines IEC 61000-4-11
70% U
(30% dip in U
) for
T
T
25 cycles
<5% U
(>95% dip in U
) for
T
T
5 sec
Power frequency (50/
3 A/m
60 Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration - electromagnetic immunity -
WSK-1011 is intended for use in the electromagnetic environment specified below. The customer or the user of
WSK-1011 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Compliance
Electromagnetic environment - guidance
level
level
Portable and mobile RF communications equipment should be used
no closer to any part of WSK-1011, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Compliance level
Electromagnetic environment - guidance
±6 kV contact
Floors should be wood, concrete or ceramic
±8 kV air
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
N/A
Note: Operate by DC Power.
N/A
Note: Operate by DC Power.
N/A
Note: Operate by DC Power.
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
Conducted RF
3 Vrms, 150
3 V
IEC 61000-4-6
kHz to 80 MHz
Radiated RF IEC
3 V/m, 80 MHz
3 V/m
61000-4-3
to 2.5 GHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radios broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which WSK-1011 is used exceeds the applicable RF
compliance level above, WSK-1011 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating WSK-1011.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
Recommended separation distances between portable and mobile RF communications equipment and WSK-1011 -
WSK-1011 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of WSK-1011 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and WSK-1011 as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter, m
Rated maximum output
150 kHz to 80 MHz,
power of transmitter, W
d=[3.5/V1]√P
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
-15-
Recommended separation distance
d=[3.5/V1]√P
d=[3.5/E1]√P, 80MHz-800MHz
d=[7/E1]√P, 800MHz-2.5GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,a should be less than the compliance
level in each frequency range.b
Interference may occur in the vicinity of equipment marked
with the following symbol:
80 MHz to 800 MHz,
800 MHz to 2.5 GHz,
d=[3.5/E1]√P
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
d=[7/V1]√P
0.23
0.73
2.3
7.3
23

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis