1.1 Transport and Storage
Please keep the original packaging for later transport, e.g. for calibration. Any transport
damage due to faulty packaging will be excluded from warranty claims.
In order to avoid instrument damage, it is advised to remove batteries when not
using the instrument over a certain time period. However, should the instrument be
contaminated by leaking battery cells, you are kindly requested to return it ti the
factory for cleaning and inspection.
Instruments must be stored in dry and closed areas. In the case of an instrument being
transported in extreme temperatures, a recovery time of minimum 2 hours is required prior
to instrument operation.
2.0 Safety Measures
The instruments Elma 1500B have been constructed and verified in compliance with the
safety measures for voltage testers DIN EN 61243-3, DIN VDE 0682 Part 401 (previously
DIN VDE 0680 Part 5), EN 61010, and IEC 1010 and have left the factory in safe and
perfect condition.
Elma 107
In order to avoid electrical shock, the valid safety and VDE
regulations regarding excessive contact voltages must
receive utmost attention, when working with voltages
exceeding 120V (60V) DC or 50V (25V)rms AC. The values
in brackets are valid forlimited ranges (as for example
medicine and agriculture).
Prior to measurement ensure that the test leads and the test
instrument are in perfect condition.
When using this instrument only the handles of the probes may
be touched – do not touch the probe tips.
The test instrument may only be used this instrument may only be
used within the ranges specified and within low voltage systems
up to 690V.
Prior to usage ensure perfect instrument function (e.g. on known
voltage source).
The voltage testers may no longer be used if one or several
functions fail or if no functionality is indicated.
Page 36