How to Switch the system ON and OFF
A. Battery status
Battery OK
Switch OFF the LED Pen when not in use to avoid draining the battery.
MVP ICON
Before completing performance verification or calibration readings:
• Ensure the LED Pen has NOT expired.
• Ensure that the battery is still functional when starting Luminescent verification or calibration (see "How to
check the battery")
If a given test point result or MVP ICON
is expected, it is suggested to confirm this result with the LED Pen.
Luminescent Verification
1. With the instrument in ATP mode, select Device >> Calibration >> Verification.
2. Select Luminescent rod.
3. Switch ON the LED Pen, check that battery LED (see "How to check battery") is lighting, then insert LED Pen
in the MVP ICON, close the door and select Verify.
4. When reading is completed, remove the LED pen and switch it OFF
ATP Performance verification
®
ATP Positive Control verification (Cat. No. 64001BC) reading is not what
®
• Slide the switch to the Right to switch ON the
• Slide the switch to the Left to switch OFF the
Replace battery*
*Once the Red Icon shows up,
please change the battery.
system ("I" marking shall be visible)
system ("o" marking shall be visible)
No Battery, Battery empty or
Battery incorrectly inserted