Chapter II
3.8.2. MAINS ADAPTER
„ Overall dimensions (L x W x H) of the housing: 96mm x 58mm x 86mm.
„ Overall dimensions (L x W x H) of the part replacing the battery compartment cover:
80mm x 37mm x 30mm.
„ Weight: approximately 450g.
„ Length of the cable: approximately 2m.
Index of protection:
IP20 as per IEC 60529 (Ed. 92)
IK04 as per IEC 50102 (Ed. 95)

3.9. COMPLIANCE WITH INTERNATIONAL STANDARDS

The device complies with the EMC and low-voltage directives required for the CE marking and with
product standard IEC 61326-1 (Ed. 2006).
3.9.1. CLAMP
Electrical safety as per IEC 61010-2-032 (Ed. 2 of 2003) for type A sensors.
Assigned characteristics:
3.9.2. MAINS ADAPTER
Electrical safety as per IEC 61010-1 (Ed. 2 of 2001).
Assigned characteristics: measurement category III, 600V referred to earth.
Oscilloscope current probe
measurement category III, 600V referred to earth.
measurement category IV, 300V referred to earth.