Err no.:
Type:
Usertext:
9149
E
When testing for leakage during the water intrusion test, the water intrusion was too high: - No filter connected - No filter
cartridge in filter housing
9150
E
The test program contains an invalid combination regarding the use of an external pressure sensor and an external valve -
Please enter the correct combination
9151
E
During the water intrusion test, a pressure increase occurred that may be caused by an increase in temperature -Please
wait for temperature compensation -Check the temperature of the wetting liquid
9152 ...
E
Internal error
9154
9155
E
The required stabilization time is too short
9156
E
During calibration/adjustment of the external pressure sensor input, the difference between two reference currents is too
low so that calibration is not possible.
9157
E
During calibration of the external pressure sensor input, all pressure values/flow rates measured were identical so that cali-
bration is not possible
9158
E
The time it takes to build up the pressure (to 100% nominal pressure) during external filling exceeds the limit
9159
E
At the end of external filling, the actual pressure in the filter housing deviates too much from the nominal pressure
9160
E
The duration of pressurization (up to 75% target pressure) during external filling exceeds the limit - The filter cartridge
may be damaged - Connect pressure supply to external valves
9161
E
The duration of pressurization (up to 75% target pressure / filter fully wet)) during external filling exceeds the limit - Con-
nect pressure supply to external valves
9162
E
During the bubble point and/or multipoint diffusion test, more than 200 pressure levels should be checked (e.g. [max. BP-
min. BP] / pressure level > 200)
9163
E
Timeout for pressure drop exceeds the limit
9164
E
The external pressure sensor indicates too high a pressure value (possible short-circuit)
9165
E
Wrong code entered so access to service mode denied
9166
E
The required test was not activated; the unit is in the service mode
9167
E
The required service function was not activated; the unit is in the test mode
9168
E
The nominal pressure decreased during filling (75% nominal pressure > nominal pressure + water column) - Do not change
pressure conversion vessel during filling - Reduce the "WIT filling factor" in the service menu
9169
E
The external pressure sensor has line defect or it is not connected
9170
E
The test has been discarded retrospectively as the external pressure sensor is not connected any longer
9171
E
During stabilization, nominal pressure exceeded target pressure
9172
E
During external filling, nominal pressure exceeded target pressure
9173
E
The flow measured is out of tolerance - select the required current and impress - if necessary, load factory-set adjustment
first
9174
E
The ambient pressure measured is out of tolerance Warning: system pressure is probably still on!
9175
E
The test was cancelled by user/operator.
9176
E
The test was cancelled by the PLC.
9177
E
The test was cancelled by switching off the power supply.
9178
E
During the MPD test, diffusion exceeded limit A2 -Please check the system for leakage -Check configuration -Rewet filter
9179
E
Wrong parameters for the MDC test.
9180
E
The differential between external and internal pressure sensor is too large.
9181
E
The diffusion at the first bubble point pressure increment exceeds the limit - Check the system for leakage - Enter the cor-
rect test class - Adjust the diffusion limit
9182
E
The pressure drop in the bubble point pressure stage exceeds the limit - Check system for leakage - Enter the correct test
class - Adjust the diffusion limit
9183 ...
E
Internal error
9187
9188
E
During volume measurement with an external reference tank, the net volume measured was so large that it was not possi-
ble to execute the test
9189
E
During volume measurement without an external reference tank, the net volume measured was so large that it was not
possible to test the filter. It is recommened that you use an external reference tank.
9190
E
Volume measurement yielded a net volume greater than 150L
9191
W
Value below a defined min. limit - No filter/filter housing connected - Filter/filter housing incorrectly connected - Check
configuration
Troubleshooting
77