4.0 TROUBLESHOOTING

4.1 Troubleshooting Table

If the flow meter fails to function, consult your provider or Maxtec.
PROBLEM
Will not shut off
Sticking float ball
Unable to set
desired flow
Knob will not turn

5.0 RECOMMENDED MAINTENANCE

5.1 Cleaning Instructions

1.
Disconnect all connections before cleaning.
2. Clean exterior surfaces of the flow meter with a cloth dampened with
a mild detergent and water.
3. Wipe dry with a clean cloth.
Caution: Do Not Autoclave
CAUSE
Leak
Defective
valve
Debris in
flow tube
Blocked inlet
Valve seized
866.4.Maxtec www.maxtec.com
REMEDY
Replace seals
and/or fittings
Replace valve
Clean flow tube
& float ball
Replace filter
Replace valve
5