2.0 SPECIFICATIONS
2.1 Flow Range Accuracy Chart
FLOW RANGE
0-30 LPM
0-70 LPM
0-120 LPM
For the most current manual revision please visit our website:
www.maxtec.com
Flush Flow is the output of the flow meter when the flow indicator is
beyond the highest calibrated graduation. The Flush Flow range is as
indicated on the flow meter labeling.
2.2 Transport/Storage Requirements
-40 °F (-40 °C) to 140 °F (60 °C)
The gas and inlet pressures are indicated on the flow tube or flow
meter body.
NOTE: Storage/transport outside the specified range may cause damage
to the flow meter.
The above flow meter models are calibrated at specified inlet pressure,
70°F (21°C), standard atmospheric pressure. Specifications are subject to
change without prior notice.
GRADUATIONS
1 (0-30) LPM
5 (0-70) LPM
10 (0-120) LPM
866.4.Maxtec www.maxtec.com
ACCURACY
2-4: ± .50 l/min,
5-30: ± 10% of
indicated value
± 10% of indicated value
± 10% of indicated value
3