Control Solution Test
Chapter 5
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label or single foil pack for exact range.
Control Range LV1
Uric Acid Test Strip
BK-U1
Self-Testing
5.8-8.6
7021
LV1
ontrol Range
5.8-8.6
mg/dL
0.34-0.51
0.34-0.51
mmol/L
General Life Biotechnology Co.,LTD.
5F., No. 240, Shinshu Rd., Shin Juang Dist.,
New Taipei City 242, Taiwan
Perform a Control Test:
Insert an Unused Strip → Make Sure the Code Number is Correct → Press "M" Button
(3 secs) → Enter Control Solution Mode (Shown CL on Screen) → Shake the Control
Solution Well → Discard First Three Drops → Put One Drop onto a Clean Surface →
Touch the Control Solution with Strip Sample Inlet → Test Result Display after Count
Down → Compare the result with the control range on test strip vial or single foil
pack
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.