MEASURiNG PROCEDURE

After selecting the appropriate material-group the unit must be hold up in the
air for automatic 0-correction.
The unit is ready for measuring after about 2 sec.
Hold the spring electrodes without excessive
force and in a approx. 40
material.
All of the 3 measuring springs must be in
good contact with the material to be measured.
Measurements may be taken in different spots
of the material, or the measuring springs could
be slided across the material.
AttENtiON:
In principle, care has to be taken regarding the base underneath. No metal
parts should be near. The best measuring results are obtained, when the mate-
rial to be measured is hold up in the air, or a Styrofoam base with a minimum
thickness of 50 mm is being put beneath the material to be measured.
MEASUREMENtS tAkEN ON PAPER:
A metal base or metal parts in the surroundings cause unfavourable influences
to the measuring results. Care has to be taken that air gaps between the layers
are avoided as much as possible. For measurements taken on turning rolls,
these should be taken after the copper brushes; just to avoid electrostatic
discharges over the unit.

BACkliGHt

To increase battery life, the backlight dims down to OFF automatically after
about 5 seconds if no measuring value changes. Immediate reactivation of
backlight after key-press or changing of measuring value.