The thermometer are intended for use in the electromagnetic environment specified
below. The customers or the users of these thermometer should assure that it is
used in such environment.
IEC 60601
Immunity test
3V
0.15 MHz to
80MHz
Conducted RF
6 V in ISM and
IEC 61000-
amateur radio
4-6
bands between
0.15 MHz and
80 MHz
Radiated RF
10V/m
IEC 61000-
80 MHz to 2.7
4-3
GHz
126
EN
Table 3
Declaration - electromagnetic immunity
Compliance
Test level
level
Not
applicable
10 V/m
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
device, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.2 P 150 KHz to 80 MHz
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 80 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
b
vicinity of equipment marked with
the following symbol: