DE/GB
GB
Instruction manual
Infrared thermometer TM A68 (model: AOJ-20M)
1
0 q w e
2
3
9
4
r
8
5
t
6
7
REF
REF
Reference number
Reference number
z
u
REF
LOT
Reference number
LOT
Batch number
Device and controls
Batch number
1
Sensor cover (remove for ear measurement; put on for forehead and direct measurement)
Sensor
Button ON/MEASURE
Backlight LCD
LED Indicator
2
3
4
5
6
Age group button (Forehead mode)
I
7
Mode button
I
LOT
se /
Previous temperature value
Current temperature value
8
9
Batch number
ON
ON
Adult forehead mode (over 12 years)
Child forehead mode
0
q
Ear mode
Direct mode
Signal tone on/off
w
e
r
Low battery indicator
Measurement unit (°C/°F)
t
z
I
Battery compartment cover
u
2 /
OFF
ON
OFF
Intended use
The TM A68 infrared thermometer from medisana is designed for measuring human body tem-
perature. The forehead measurement mode is suitable for use by people of all ages and the ear
measurement mode for use by people over three months of age. Additionally, the temperature of
objects can be measured under direct mode.
Dispose of packaging in an
OFF
Dispose of packaging in an
Contraindication
environmentally friendly manner
environmentally friendly manner
Do not use the thermometer if the ear is infected with otitis or suppuration.
Explanation of symbols
Dispose of packaging in an
IMPORTANT!
environmentally friendly manner
雨
Refer to instruction manual recommended.
CAUTION /
Importer
Importer
注意!查阅随机文件
WARNING
These warning notes must be observed to prevent any injury to the user.
Type BF applied part /
CAUTION
BF型应用设备
These notes must be observed to prevent any damage to the device.
Importer
NOTE
Read the instructions for use /
These notes give you useful additional information on the installation or operation.
请阅读说明书
CAUTION /
REF
Reference number
注意!查阅随机文件
Indication of the degree of protection against foreign bodies and water
Device in protection class 2 /
IP22
against solid foreign bodies of 12.5 mm Ø and more / When the thermometer is held at an angle of
Type BF applied part /
LOT
二类设备
15 degrees, no water droplets penetrate)
Batch number
BF型应用设备
port
Classification: Type BF applied part
LOT number
I
Read the instructions for use /
ON
Manufacturer / 制造商
请阅读说明书
Manufacturer
Date of manufacture
Device in protection class 2 /
This product must not be disposed
OFF
Unique device identifier
二类设备
ture
of together with domestic waste.
Keep dry / 保持干燥 / 怕雨
Dispose of packaging in an
Use by
Manufacturer / 制造商
CE marking
environmentally friendly manner
0123
WEEE
21
21
Recycling symbols/codes: These are used to provide information about the material
Keep dry / 保持干燥 / 怕雨
and its proper use and recycling.
Importer
04
20
Storage/Transport
04
04
20
20
Storage/Transport
SN
Serial number
PAP
PAP
PAP
WEEE
70°C
95%
1060hPa
70°C
95%
1060hPa
For France: Our products and packaging are recycled, don't throw them away! You can
REF
Reference number
find appropriate drop-off points on the website „www.quefairedemesdechets.fr".
-25°C
10%
700hPa
Storage/Transport
10%
700hPa
-25°C
Permissible storage and transport
SN
Serial number
temperature and humidity
LOT
Storage/Transport
Batch number
Operating
Operating
Storage/Transport
I
70°C
Permissible storage and transport
95%
1060hPa
1060hPa
40°C
95%
Operation
95%
1060hPa
temperature and humidity
40°C
ON
Permissible operating temperature
Temperature range
Humidity range
Ambient pressure limitation
and humidity
-25°C
10%
700hPa
10°C
700hPa
10%
10°C
700hPa
10%
Operation
OFF
Permissible operating temperature
Authorized Representative
and humidity
EC
REP
Operating
In The European Community
SN
Serial number
Medical device
Dispose of packaging in an
95%
1060hPa
40°C
environmentally friendly manner
Medical device
Medical device
Protect from moisture
Storage/Transport
10°C
700hPa
10%
70°C
95%
1060hPa
Importer
Distributor
Importer
-25°C
Distributor
10%
700hPa
Distributor
CAUTION /
Operating
注意!查阅随机文件
40°C
95%
1060hPa
10°C
700hPa
Keep away from sunlight
Type BF applied part /
This way up
10%
Date of manufacture
Date of manufacture
BF型应用设备
Authorised representative of
Read the instructions for use /
Fragile, handle with care
e
请阅读说明书
CH
EC
REP
Switzerland
Do not use outdoors
(indoor use only)
Device in protection class 2 /
Do not use outdoors
二类设备
(indoor use only)
Shenzhen AOJ Medical Technology Co., Ltd.
Manufacturer / 制造商
Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Com-
Single patient multiple use
munity, Xixiang Street, Bao'an District, 518126 Shenzhen, CHINA
Storage/Transport
Single patient multiple use
Keep dry / 保持干燥 / 怕雨
70°C
95%
1060hPa
PA
Share Info GmbH
EC
REP
中国GCC血压计计量证号
-25°C
10%
700hPa
Heerdter Lohweg 83, 40549 Düsseldorf, Germany
2020F213-44
Operating
WEEE
Tel: 0049 179 5666 508; E-mail: EU-Rep@share-info.com
40°C
95%
1060hPa
PA
10°C
700hPa
中国GCC血压计计量证号
10%
SN
Serial number
2020F213-44
Storage/Transport
Permissible storage and transport
temperature and humidity
Operation
Permissible operating temperature
and humidity
1
PL/FI HU BE/NI/IE
IAN 470270_2404
Medical device
RO/CY
8
Distributor
GB Safety Information
Read the instruction manual carefully before using this device, especially the safety in-
structions, and keep the instruction manual for future use. Should you give this device
to another person, it is vital that you also pass on these instructions for use.
• Only use the unit for the specific purpose described in the instruction manual. Any misuse will void the warranty.
• The unit is not designed for commercial use.
• Keep the appliance away from children under 12 years of age. When using on children, they must be supervised by
an adult. Please observe the general safety instructions when using on children.
• Never immerse the thermometer in water or other liquids, it is not waterproof. For cleaning and care, please follow
the instructions in the corresponding chapter in this manual.
• The sensor must be cleaned after each use (see chapter „Cleaning and care").
• Keep the thermometer away from direct sunlight and store it in a dust-free, dry and well-ventilated place in accord-
ance with the „Storage conditions" in the chapter „Technical data". Only use the thermometer in environments as
specified in „Operating conditions".
• Do not touch the tip of the temperature sensor - it is very sensitive.
• Always keep the temperature sensor clean to ensure accurate readings.
• Do not use the thermometer if the sensor or the unit itself show signs of damage. If the unit is damaged, do not at-
tempt to repair it! Please contact customer service.
• No modifications are permitted on this thermometer. Unauthorised modification may result in serious hazards.
• This thermometer is made of high quality precision parts. Do not drop the device. Protect it from strong shocks and
impacts. Do not twist the device or the sensor and generally handle the device with great care.
• Contact your doctor if you notice symptoms such as unexplained irritability, vomiting, diarrhoea, dehydration, chang-
es in appetite or activity, seizures, muscle pain, chills, stiff neck, pain when urinating, etc. - even if there is no fever.
• Even people with normal body temperature (if no fever is detected) may need medical attention. For people taking
antibiotics, painkillers or antipyretics, the severity of the illness should not be judged by temperature alone.
• Elevated fever can be an indication of serious illness, especially in elderly, frail or immunocompromised adults and
in newborns and infants. Please seek professional advice immediately if an increase in temperature occurs when
the following groups of people are measured: - people over 60 years of age (fever may be attenuated or absent in
elderly patients), - people with diabetes mellitus or weakened immune system (e.g. HIV positive, cancer, chemo-
therapy, chronic steroid treatment, splenectomy), - bedridden (e.g. nursing home patient, stroke, chronic illness),
- transplant patients (e.g. transplants of liver, heart, lung, kidney).
• This thermometer is not suitable for premature or gestational age babies. This thermometer is not suitable for inter-
preting hypothermic temperatures. Do not leave children unattended to take their temperature.
• Do not use the thermometer on newborns or for continuous temperature monitoring.
• Do not take a reading during or immediately after breastfeeding a baby.
• Patients should not drink, eat or exercise before or during the measurement.
• The use of this thermometer is not a substitute for a consultation with your doctor or paediatrician.
• Do not take measurements in quick succession, always wait at least 10 seconds.
• The thermometer is not intended to diagnose or treat any health problem or disease. The measurement results are
for reference only, always consult your doctor.
BATTERY SAFETY INSTRUCTIONS
• Do not disassemble the batteries!
• Remove exhausted batteries from the unit immediately! Also remove the batteries if you do not use the unit for 2
months or longer!
• Do not use rechargeable batteries!
• Keep the battery compartment closed!
• Increased risk of leakage, avoid contact with skin, eyes and mucous membranes! In case of contact with battery
acid, rinse the affected areas immediately with plenty of clean water and consult a doctor immediately!
• If a battery has been swallowed, consult a doctor immediately!
• Insert the batteries correctly, observe the polarity!
REF
• Only insert batteries of the same type, do not use different types or used and new batteries together!
Reference number
• Keep batteries away from children!
• Do not recharge batteries! Do not short-circuit! Do not throw into fire! Danger of explosion!
• Store unused batteries in the packaging and not near metallic objects to avoid short-circuiting!
LOT
Batch number
Items supplied and packaging
Please check first of all that the device is complete and is not damaged in any way. In case of doubt, do not use it and
I
contact your dealer or the service centre.
The following parts are included:
ON
• 1 medisana Infrared thermometer TM A68
• 1 Pouch
(Protected
• 2 Batteries 1.5V (type AAA/LR03 )
OFF
• Instruction manual
The packaging can be reused or recycled. Please dispose properly of any packaging material no longer required. If
you notice any transport damage during unpacking, please contact your dealer without delay.
Dispose of packaging in an
WARNING - Pay attention, that the polythene packing is kept away from the reach of children!
environmentally friendly manner
Risk of suffocation!
Product description
The infrared thermometer measures body temperature based on the infrared energy emitted from the eardrum or fore-
Importer
head. After correctly positioning the temperature sensor in the ear canal or on the forehead, the user can quickly obtain
measurement results. Normal body temperature lies within a certain range. The following table shows that this normal
range also varies from measurement location to measurement location. Therefore, readings from different locations
should not be directly compared. Tell your doctor what type of thermometer you used to take your temperature and at
which body site. Keep this in mind even if you are self-diagnosing.
Measuring site
Forehead
36.1°C - 37.5°C (97°F - 99.5°F)
Ear
35.8°C - 38°C (96.4°F - 100.4°F)
Oral
35.5°C - 37.5°C (95.9°F - 99.5°F)
Rectal
36.6°C - 38°C (97.9°F - 100.4°F)
Armpit
34.7°C - 37.3°C (94.5°F - 99.1°F)
Features
• The option of selecting an age group (0 - 6 years / 6 -12 years / over 12 years) enables more accurate measurement
results. For the child age groups up to 12 years, we recommend measurement in forehead mode.
• The device can store up to 40 measurement results and has a silent and a signal tone mode as well as an acoustic
Storage/Transport
and optical fever warning.
70°C
95%
1060hPa
Remove / insert batteries
-25°C
10%
700hPa
If the low battery indicator
t
or nothing at all appears in the LCD display
Operating
REF
1.5V batteries (type AAA / LR03) must be inserted. Note: When using the unit for the first time, the insulating strip
Reference number
95%
1060hPa
40°C
may need to be removed for the unit to work.
10°C
700hPa
LOT
Changing the batteries: Open the battery compartment cover
10%
Batch number
batteries (type AAA / LR03). Make sure that the batteries are lying in the correct direction. Close the battery compart-
I
ment. When the batteries are changed, all stored results are deleted.
ON
Operation - Notes on the measurements
OFF
To avoid inaccurate results, please note the following points:
• Only use indoors where there is no wind.
Dispose of packaging in an
environmentally friendly manner
• Measure only when you are relaxed. Previous, strong movements (e.g. sport) or strong tension can lead to incor-
rect readings.
• The ambient temperature must not be extremely high or low. To ensure accurate readings, keep the thermometer
Importer
at room temperature for more than 30 minutes before use.
• Do not hold the thermometer in your hand for a long time as it is very sensitive to heat.
medisana GmbH
• The device has been subjected to clinical tests. Safe use can be achieved by following the instructions given in
Carl-Schurz-Str. 2,
this manual.
41460 NEUSS, Germany
0123
Storage/Transport
1060hPa
70°C
95%
10%
700hPa
-25°C
Operating
40°C
95%
1060hPa
10°C
700hPa
Take a temperature on the forehead
1.
Make sure that the sensor cover
2.
Use button
6
to set the age group (0 - 6 years / 6 -12 years / over 12 years).
3.
If not already preselected automatically, select the measuring mode
forehead measurement) using the mode button
4.
Aim the thermometer at the centre of the forehead to measure the body temperature and keep a distance of 1 - 3
cm. Press and hold button
signal and displays the measured value.
5.
Without further input, the unit switches off after approx. 13 seconds.
Taking a temperature in the ear
1.
Remove the sensor cover
1
2.
If not already automatically preselected, select the measurement mode
button
.
7
3.
Carefully insert the sensor
temperature. For children up to 1 year of age, it is advisable to pull the ear back slightly when doing
this (see fig. a). For older children and adults, the ear can be pulled slightly upwards and backwards
(see fig. b) to bring the ear canal into a straight course so that the sensor can take a temperature
reading at the eardrum.
4.
Press and hold button
for approx. 1 second. The device measures the temperature, emits an
3
acoustic signal and displays the measured value. Now carefully remove the thermometer from the
ear and release the ear again.
5.
Without further input, the unit switches off after approx. 13 seconds.
WARNING! Prevent injury or inaccurate measurement results:
• Remove earwax or blockages in the ear canal before taking a measurement.
• Never force the sensor into the ear canal.
• The measurement in the right ear may differ from the measurement in the left ear.
• Therefore, always measure the temperature in the same ear.
• Be especially careful with children because of their small ear canal.
• We recommend forehead measurement for children up to 12 years.
Measuring an object temperature
1.
Make sure that the sensor cover
2.
Select the measurement mode
3.
Aim the thermometer at the centre of the object and keep a distance of 1 - 3 cm. Press and hold button
approx. 1 second. The unit measures the temperature, emits an acoustic signal and displays the measured value.
4.
Without further input, the unit switches off after approx. 13 seconds.
Classification of the measured temperature (T)
Ear or forehead measurement: Depending on the measured temperature, the display appears with different back-
lighting or different signal tones are emitted (not in silent mode), see the following list. In direct mode, a long tone
always sounds and the display is backlit in white.
32°C (89.6°F) ≤ T ≤ 37.5°C (99.5°F) : Illumination green, one long beep
37.6°C (99.7°F) ≤ T ≤ 38°C (100.4°F) : Illumination orange, 6 short beeps (slight fever).
38.1°C (100.6°F) ≤ T ≤ 42.9°C (109.2°F) : Illumination red, 6 short signal tones (high fever)
Switch signal tones on/off
You can switch the signal tones on or off. To do this, press and hold the button
is switched on. Depending on the currently set status, either the corresponding symbol
or the symbol disappears (signal tones on).
Set measuring unit
The measuring unit can be set to degrees Celsius (°C) or degrees Fahrenheit (°F). To change the unit, press and
hold button
6
for approx. 2 seconds when the unit is switched on.
Retrieving / deleting memory data
The measured temperatures are stored by the unit (max. 40 results). With the unit switched on, press and hold the
mode button
7
for approx. 5 seconds to call up the values in the memory. The last result measured is displayed first.
You can also view the values stored before by pressing the mode button
Important notes regarding the determination of body temperature
• It is important to know the normal temperature of each person when they are healthy. This is the only way to ac-
curately diagnose a fever. Take the temperature twice a day (early morning and late afternoon). Take the average
of the two temperatures to calculate the normal oral equivalent temperature. Always take the temperature in the
same place, as temperature readings may be different in different places.
• A child's normal temperature can vary between 36.1°C (97ºF) and 37.7°C (99.9°F). Please note that this device
reads 0.5ºC (0.9ºF) lower than a rectal digital thermometer.
• External factors can noticeably affect the temperature reading, such as: - prolonged lying on one ear, - covering
with hair, - prolonged exposure to high or low ambient temperatures, - bathing or swimming before taking a read-
ing. You should always relax for about 20 minutes before the measurement in a well-tempered, quiet environment
so as not to falsify the result. If you want to take an ear measurement and use medication (e.g. ear drops) in one
ear, use the other ear for the measurement.
• Always remove hair, sweat, dirt, etc. before taking a measurement and wait 10 minutes before taking the measure-
ment.
• In the following situations, it is recommended to take 3-5 measurements at the same site and take the highest
value as the reading: • In newborn children in the first 100 days, • In children under 3 years of age with a weakened
immune system where the presence or absence of fever is critical, • For first-time use, until the user has become
familiar with the device and is getting consistent readings.
Cleaning and care
Clean the sensor
at least 5 minutes before and after each use. You can use a cotton swab
2
moistened with 70% alcohol for this. After the alcohol has dried completely, you can take a
Normal body temperature
new measurement. After use, the housing should also be cleaned with a soft, dry or very
slightly damp cloth. Always make sure that no liquid gets inside the thermometer. Never use
abrasive cleaners, thinners or benzene for cleaning and do not immerse the device in water
or other cleaning liquids. Be careful not to scratch the surface of the LCD display.
Troubleshooting
Problem: The thermometer fails to power on.
Possible cause: Low batteries / wrong polarity / thermometer is damaged.
Solution: Exchange batteries and make sure, these are installed correctly.
Problem: "Er1" is displayed.
Possible cause: The ambient temperature is lower than 0°C (32°F) or higher than 40°C (104°F).
Solution: Take a measurement in the range of the operating conditions.
Problem: "ErC" is displayed.
Possible cause: Memory error or temperature correction not completed.
4
after a certain period of operation, 2 new
Solution: Thermometer switches off after 3 seconds. Then repeat the measurement!
Problem: "HI" or "LO" is displayed with red backlight.
u
and remove the batteries. Insert two new 1.5 V
Possible cause: Measured temperature too high (above 42.9°C/109.2°F [ear/forehead] / 100°C/199°F [direct] or too
low (below 32°C/89.6°F [ear/forehead] or 0°C / 32°F [direct].
Solution: Repeat the measurement!
Problem: The temperature reading is lower than the typical body temperature range.
Possible cause: The lens of the temperature sensor is dirty / thermometer sensor is not aligned to eardrum / the ther-
mometer is used within 30 minutes after being taken from a cold environment.
Solution: Clean the lens using a cotton swab / reposition the thermometer probe so that it is aligned to the eardrum /
wait for more than 30 minutes after the device is moved into the measurement environment.
Problem: The temperature reading is higher than the typical body temperature range.
Possible cause: The temperature sensor is damaged or you are coming from a hot environment.
Solution: Wait more than 30 minutes after bringing the unit into the measurement environment.
If the problems cannot be solved with these instructions or other problems occur, please contact the customer ser-
vice.
Disposal
is in place and switch on the thermometer by briefly pressing the button
.
1
3
0
(adult forehead measurement) or
q
(child
.
7
household waste, but at a battery collection station at a recycling site or in a shop. Consult your municipal authority
3
for approx. 1 second. The device measures the temperature, emits an acoustic
or your supplier for information about disposal.
Technical specifications
Name and model:
Standards applied:
from the device and switch on the thermometer by briefly pressing the button
.
3
Display system:
w
(ear measurement) using the mode
Power supply:
Measurement unit:
a
2
into the ear canal without applying great force to measure the body
Measuring range:
Measurement accuracy:
(Laboratory readings)
Display resolution:
Automatic switch-off:
Memory capacity:
Operating conditions:
b
Storage conditions:
Product life:
Batteries:
Dimensions:
Weight:
Article No. / EAN Code:
1
is in place and switch on the thermometer by briefly pressing the button
3
.
(direct mode) using the mode button
.
e
7
Guarantee and repair conditions
3
for
medisana GmbH guarantee: Dear Customer, a guarantee on this device is provided which is valid for 3 years from the
date of purchase. Should this product be defective, you can assert your legal rights against the retailer. These legal
rights are not affected by our guarantee, which is described in the following.
Processing a guarantee claim
To ensure your request is processed quickly, please observe the following information:
1. Should functional errors or other faults occur, please start by contacting the service point specified
in the following by telephone or by e-mail.
2. Please keep your receipt and the item number IAN 470270_2404 on hand as proof of purchase.
3. A product which has been registered as defective can be sent, without postage, to the Service
address given to you by the Service point. You must enclose proof of purchase (receipt) and details
about the defect and when it occurred.
3
for about 5 seconds while the unit
Guarantee conditions
r
appears (signal tones off)
The guarantee period begins with the date of purchase. Please keep the original sales receipt. This document is
required for proof of purchase. If a material or production fault occurs in this product within 3 years of the date of
purchase, the product will be repaired for you free of charge, or replaced – at our discretion. This guarantee service
requires that the defective device and the proof of purchase (receipt) are sent in, along with a brief description of
what the defect is and when it occurred, within the three-year period. If the defect is covered by our guarantee, the
product will be returned to you as a repaired product or a new product. No new guarantee period begins upon repair
or replacement. Any damage and defects already present at the time of purchase must be reported immediately after
unpacking. Once the guarantee period has expired, you will be billed for any repairs required. The guarantee service
applies for material or production faults. This guarantee does not extend to product components subject to normal
. If no other value exists, "---" is displayed.
7
wear and can therefore be considered to be wearing parts or damage to fragile parts, e. g. switches or parts made of
glass. This guarantee expires if the product is damaged, has not been used properly, or was subject to maintenance
work. To ensure the product is used properly, all instructions included in the operating manual must be complied with
exactly. Using the product for purposes or procedures which the operating manual discourages or warns against
must be avoided. The product is intended for private use only and not for commercial use. If the product is subject
to improper and inappropriate treatment, the use of force, or service work which is not carried out by our authorised
service office, the guarantee will be voided.
Service
GB /
Service United Kingdom
phone: 00800 63347262; eMail: service.gb@medisana.com
IE
Service Republic of Ireland
phone: 00800 63347262; eMail: service.ie@medisana.com
This product must not be disposed of together with domestic waste. All users are obliged to hand in all electri-
cal or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or com-
mercial collection point so that they can be disposed of in an environmentally acceptable manner.
Please remove the batteries before disposing of the device/unit. Do not dispose of old batteries with your
medisana Infrared thermometer TM A68 (model: AOJ-20M)
ASTM E1965-98 / ISO80601-2-56
Digital display (LCD), 4-colour backlight (white,green,orange,red)
3V , 2 x AAA batteries 1.5V
°C/°F, switchable
Forehead/ear measurement: 32.0°C~42.9°C (89.6°F~109.2°F)
Direct measurement: 0.0°C~100.0°C (32°F~199°F)
Forehead/ear measurement: ±0.2°C (±0.4°F);
Direct measurement: ±1.0°C/±2.0°F
0.1°C/0.1°F
After 13±1 seconds
Up to 40 measurement results
Temperature 10°C-40°C (41°F-104°F), rel. humidity:15-95% non-
condensing, atmospheric pressure: 70-106 kPa
Temperature -20°C-55°C (-4°F-131°F), rel. humidity:0-95% non-
condensing, atmospheric pressure: 50-106 kPa
5 years
2xAAA, can be used more than 3000 times
approx. 162 x 43 x 35 mm
approx. 76 g without batteries
92139 (IAN 470270_2404) / 40 15588 92139 1
In accordance with our policy of continual product improvement, we reserve the right to make technical and
optical changes without notice.
IAN 470270_2404