Electromagnetic immunity
The Naída CI is intended for use in the electromagnetic environment specified
below. The customer or the user of the Naída CI should assure that it is used in such
an environment.
Immunity
IEC 60601
test level
Electrostatic
± 6 kV
discharge (ESD)
contact
IEC 61000-4-2
± 8 kV air
Power frequency
3 A/m
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
26
Compliance
Electromagnetic environment - guidance
level
a
± 6 kV
Floors should be wood, concrete or ceramic tile. If
contact
floors are covered with synthetic material, the relative
± 8 kV air
humidity should be at least 30 %. As with the operation
of other electronic devices, precaution should be taken
to not generate ESD.
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.