To replace the batteries unscrew the battery
compartment cap and replace all four 1.5V batteries
while paying attention to their polarity. Make sure
the gasket is in place before screwing back the cap.
Batteries should only be replaced in a
nonhazardous area using the battery type specified
in this instruction manual.
ACCESSORIES
MI56P
Replaceable electrode for pH55 & pH56
M10004B
pH 4.01 buffer, 20 ml sachet, 25 pcs.
M10007B
pH 7.01 buffer, 20 ml sachet, 25 pcs.
M10010B
pH 10.01 buffer, 20 ml sachet, 25 pcs.
MA9004
pH 4.01 buffer, 230 ml bottle
MA9006
pH 6.86 buffer, 230 ml bottle
MA9007
pH 7.01 buffer, 230 ml bottle
MA9009
pH 9.18 buffer, 230 ml bottle
MA9010
pH 10.01 buffer, 230 ml bottle
MA9015
Electrode storage solution, 230 ml
MA9016
Electrode cleaning solution, 230 ml
Electrode rinse solution, 20 ml sachet,
M10000B
25 pcs.
SPECIFICATIONS
RANGE
-2.0 to 16.0 pH (pH55);
-2.00 to 16.00 pH (pH56);
-5.0 to 60.0°C / 23.0 to 140.0°F
RESOLUTION
ACCURACY (@25°C)
TYPICAL EMC DEVIATION
TEMPERATURE COMPENSATION
CALIBRATION
Automatic, 1 or 2 points
with 2 sets of memorized buffers
(pH 4.01, 7.01, 10.01 or 4.01, 6.86, 9.18)
PROBE
Replaceable MI56P
ENVIRONMENT
-5 to 50°C; 100% RH max.
BATTERY TYPE
4 x 1.5V; IEC LR44, A76
BATTERY LIFE
approx. 300 hours of use
AUTO-OFF
after 8 minutes of non-use
DIMENSIONS
WEIGHT
CERTIFICATION
Milwaukee Instruments conform to the CE
European Directives.
Disposal of Electrical & Electronic Equipment.
Do not treat this product as household waste. Hand
it over to the appropriate collection point for the
recycling of electrical and electronic equipment.
Disposal of waste batteries. This product contains
batteries. Do not dispose of them with other
household waste. Hand them over to the appropriate
collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences
for human health and the environment. For
detailed information, contact your local
household waste disposal service or go to
www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely
suitable for your specific application and for the
environment in which it is used. Any modification
introduced by the user to the supplied equipment
may compromise the meter's performance. For your
0.1 pH (pH55);
and the meter's safety do not use or store the meter in
0.01 pH (pH56);
hazardous environment. To avoid damage or burn, do
0.1°C / 0.1°F
not perform any measurement in microwave ovens.
±0.1 pH (pH55);
WARRANTY
±0.05 pH (pH56);
±0.5°C / ±1°F
These instruments are warranted against defects in
±0.1 pH (pH55);
materials and manufacturing for a period of 2 years
±0.02 pH (pH56);
from the date of purchase. Probe is warranted for
±0.3°C / ±0.6°F
6 months This warranty is limited to repair or free
of charge replacement if the instrument cannot
Automatic,
be repaired. Damage due to accidents, misuse,
from -5 to 60°C
tampering or lack of prescribed maintenance is not
covered by warranty. If service is required, contact
your local Milwaukee Instruments Technical Service.
If the repair is not covered by the warranty, you will
be notified of the charges incurred. When shipping
any meter, make sure it is properly packaged for
complete protection.
Milwaukee Instruments reserves the right to
make improvements in design, construction and
appearance of its products without advance notice.
200 x dia 38 mm
100 g
ISTPH56 07/20
USER MANUAL
pH55 & pH56
PRO Waterproof
pH/Temperature
milwaukeeinstruments.com (USA & CAN)
SATISFACTION
GUARANTEED
Testers
pH56
milwaukeeinst.com

Werbung

loading

Inhaltszusammenfassung für Milwaukee pH55

  • Seite 33 GERMAN BENUTZERHANDBUCH – pH55 & pH56 PRO Wasserdichte pH/Temperatur-Tester Um die Batterien zu ersetzen, schrauben Sie den Batteriefachdeckel ab und ersetzen Sie alle vier 1,5-V-Batterien unter Beachtung ihrer Polarität. Vergewissern Sie sich, dass die Dichtung richtig sitzt, bevor Sie den Deckel wieder zuschrauben.
  • Seite 34 Wartung sind nicht durch die Garantie abgedeckt. Wenn eine Wartung erforderlich ist, wenden Sie sich an den technischen Kundendienst von Milwaukee Instruments vor Ort. Wenn die Reparatur nicht durch die Garantie abgedeckt ist, werden Sie über die anfallenden Kosten informiert. Wenn Sie ein Messgerät versenden, stellen Sie sicher, dass es für einen vollständigen Schutz ordnungsgemäß...
  • Seite 35 Drücken Sie eine beliebige Taste, um zum normalen Modus zurückzukehren. Um das Messgerät auszuschalten, drücken Sie die Taste ON/OFF. Die Meldung „OFF“ erscheint auf der zweiten Anzeige. Lassen Sie die Taste los. Hinweis: Vergewissern Sie sich vor jeder Messung, dass das Messgerät kalibriert ist (die CAL-Anzeige leuchtet). Schalten Sie das Messgerät nach Gebrauch immer AUS, spülen Sie die Elektrode mit Wasser ab, um Verunreinigungen zu minimieren, und bewahren Sie sie mit einigen Tropfen Aufbewahrungslösung (MA9015) oder pH7-Lösung (M10007) in der Schutzkappe auf.
  • Seite 36 Um den Setup-Modus aufzurufen, drücken Sie die Taste ON/OFF, bis „CAL“ auf der zweiten LCD-Anzeige durch „TEMP“ und die aktuelle Temperatureinheit (z. B. TEMP °C) ersetzt wird. Dann: - für die Auswahl °C/°F: Verwenden Sie die Taste SET/HOLD; drücken Sie dann einmal die Taste ON/OFF, um die Auswahl des Puffersatzes aufzurufen, oder zweimal, um zum normalen Messmodus zurückzukehren.

Diese Anleitung auch für:

Ph56