Directions for Use

Braun ThermoScan Pro 4000 Model

Caution Store thermometers for testing in the same
room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to allow for
thermal accommodation.
Figure 5. 9600 Plus Calibration Tester with the Braun ThermoScan
Pro 4000
To begin calibration verifcation of the thermometers:
1. Clean the probe tip with a cotton swab slightly moistened
with isopropyl alcohol, remove excess alcohol with a clean
Chapter 3 Operation
15