BATTERY CHANGE
To ensure proper functioning, replace the battery when
the battery icon is displayed. If you are not planning to use
the device for a longer period of time, please remove the
battery.
EN
Make sure that the unit is is switched off.
Open the battery compartment (slide the
cover down).
Remove the battery.
Insert new battery (please pay attention
to polarity / suitability).
Close battery compartment.
FUNCTION AND SAFETY TEST
For safety reasons, the following precautions should be
carried out once a week.
Check the device for any external damage:
Check for deformation of the housing or damage to the
electrode output sockets.
Check that no type plates or stickers are damaged.
Check the indication of the LEDs when switching on.
Check cables and electrodes for damages.
A safety check must be performed by authorized person-
nel at least every two years by the end of the month in
which it is due.
The Instructions for use must always be carried with the
unit. If there are any problems or defects, please contact
your retailer.
24
REPAIR MINOR FAULTS YOURSELF
If the device does not work properly, the instructions below
may help identify the cause of the fault. If the problem
cannot be solved with any of these tips, contact the custom-
er service.
Check cables and connections! Have all plugs been con-
nected correctly?
Does the display not switch on? Change batteries.
If a cable is externally damaged, replace it immediately
with a good one. Only use original accessories.
No current is flowing? Check that the electrodes are cor-
rectly connected to the same channel and that the current
is strong enough.
CONFORMITY TO SAFETY STANDARDS
The STIM-PRO Comfort+ device is in compliance with the
safety standards EN 60601-1 Medical electrical equipment
- Part 1: General requirements for basic safety and
essential performance & EN 60601-1-2 Medical electrical
equipment - Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard:
Electromagnetic disturbances - Requirements and tests.
WARRANTY
All devices have a warranty of 24 months from the date
of purchase. The warranty applies to the stimulator only
and covers both parts and labor relating thereto. The
warranty does not apply to damage resulting from improper
handling, the failure to follow the instructions for use, loss
or dropping.