Prüfpume/Test pump
C1351X085005
Never actuate the change-over valve when the hand test pump
is under pressure or vacuum! Actuate the change-over valve
only when the relief valve is open.
Remove the reference instrument or the test specimen only
when the relief valve (3) is open and no vacuum is in the hand
test pump any more.
As a standard feature, the
1385.085.005
been
adjustable lift stop in order
to
overpressure
and test specimen).
The upper knurled nut is
used to set the lift stop, and
the lower knurled nut is
used to fix the setting. A
shorter lift travel permits
the
pressures
vacuum generation more
easy); a longer lift travel will
generate higher pressures.
5
Maintance instructions
Prior to connecting the reference instrument and the test specimen, the
sealing gaskets in the two conne ctors should be checked for correct
position and wear, and should be replaced, if and when necessary.
A service kit, consisting of spare sealing gaskets and o -rings, is available
as an accessory.
18
pump
fitted
with
reduce
the
risk
(reference
generation
of
(and
makes
has
an
of
lower
Overpressure protective device
BD_BE 527 a