RETURN AND DISPOSAL
This device may not be disposed of with household waste . After use of the device, please return the device to the provider for disposal . This device contains electrical and/or
electronic components that must be recycled per EU Directive 2012/19/EU-Waste Electrical and Electronic Equipment (WEEE) . Non-infectious used accessories (e .g . nasal
cannula) can be disposed of as residential waste . The disposal of infectious accessories (e .g . nasal cannula from an infected user) must be made via an approved waste
disposal company . Names and addresses can be obtained from the local municipality .
SPECIFICATIONS
Catalog Number
Delivery Rate (Lower delivery rates
available for low flow applications)
Maximum Recommended Flow (@
nominal outlet pressures of zero and
7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-4921 ft)
1500-4000 M (4921-13123 ft)
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity
range of 10% to 95%
Power Consumption
Weight
Sound Level (ISO 8359:1996)
Dimensions
Pressure Relief Valve
Operating System
The visible "low oxygen" indicator
will activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
EMC Compliance To
*NOTE– The OSD performance at 41˚F (5˚C) to 104˚F (40˚C), 95% R.H. through voltage range on the 525DS verified at 670m.
**The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
Specifications subject to change without notice.
EN - 8
DEVILBISS 5-LITER SERIES
525DS, 525DS-Q
0 .5 to 5 LPM
5 LPM
8 .5 psig (58 .6 kPa)
Outlet Pressure: 14-21 psig
(96 .5-144 .8 kPa)
Outlet Flow: 2 LPM Continuous
115 V, 60 Hz, 3 .3 Amp
97-127 V~, 60 Hz
1-5 LPM=87%-96%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation of performance
No degradation in performance across
the operating voltage range .
310 Watts Average
275 Watts @ 1 .2 LPM & below
36 lbs . (16 .3 Kilograms)
48 dBA (525DS)
46 dBA (525DS-Q)
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
40 psig±5psig (276 kPa±34.5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
(The audible alarm will alert at
approximately 75%. At less than 60%, the
red "service required" light will activate .)
-40˚F (-40˚C) to 158˚F (70˚C), humidity
range of 10% to 100%, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV
IEC 60601-1 3rd Edition
EN60601-1-2
525KS, 525KS-LT
0 .5 to 5 LPM
5 LPM
8 .5 psig (58 .6 kPa)
Outlet Pressure: 14-21 psig
(96 .5-144 .8 kPa)
Outlet Flow: 2 LPM Continuous
220-230 V~, 50 Hz, 1 .55 Amp
230 V~, 60 Hz, 1 .9 Amp
187-253 V~, 50 Hz
195-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/50Hz only:
No degradation of performance
No degradation in performance across
the operating voltage range .
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @
1 .2 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @
1 .2 LPM & below
36 lbs . (16 .3 Kilograms)
< 40 dBA (50Hz) (525KS)
48 dBA (50Hz) (525KS-LT)
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
40 psig±5psig (276 kPa±34.5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
(The audible alarm will alert at
approximately 75%. At less than 60%, the
red "service required" light will activate .)
-40˚F (-40˚C) to 158˚F (70˚C), humidity
range of 10% to 100%, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV approved for 50 Hz only to
EN 60601-1:1990+A1+A2+A13
ISO 8359: 1996/Amd .1:2012
EN60601-1-2
525PS
0 .5 to 5 LPM
5 LPM
8 .5 psig (58 .6 kPa)
N/A
220-230 V~, 60 Hz, 1 .68 Amp
187-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/60Hz only:
No degradation of performance
No degradation in performance across
the operating voltage range .
230V / 60 Hz - 334 Watts Average
230V / 60 Hz - 297 Watts Average @
2 .5 LPM & below
36 lbs . (16 .3 Kilograms)
40 dBA (50Hz)
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
40 psig±5psig (276 kPa±34.5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
(The audible alarm will alert at
approximately 75%. At less than 60%, the
red "service required" light will activate .)
-40˚F (-40˚C) to 158˚F (70˚C), humidity
range of 10% to 100%, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV
EN 60601-1:1990+A1+A2+A13
ISO 8359: 1996/Amd .1:2012
EN60601-1-2
SE-525K3

Werbung

loading

Diese Anleitung auch für:

525ds525ds-q525ks525ks-lt525ps