• Do not store in the refrigerator any explosive materials or spray cans with combustible
propellants such as propane, butane and pentane. Any escaping gases could be ignited by
electrical components. These kinds of spray cans can be recognized as such by the printed
details on content or a flame symbol
• Do not use any electrical equipment inside the refrigerator.
• Keep keys of lockable equipment well out of the reach of children.
• Only load the refrigerator with 2 - 8°C refrigerated products (intrinsic temperature).
Proper cooling cannot be ensured if this is disregarded.
• Do not put any foodstuffs/drinks into the refrigerator!
• Visually check information labels, type plate and magnetic contact at least twice a year
as to their legibility/intactness. If any damage is detected, please immediately contact
the Customer Services department.
• Do not keep any „frozen products" (cold packs/pads, ice etc.) in the refrigerator.
Proper cooling cannot be ensured if this is disregarded!
• To ensure a proper data record, date and time must always be up to date.
Please check this regularly.
The lower/upper section of the refrigerator (see drawing) must not be blocked.
The recirculation function will be disturbed and proper ventilation cannot be guaranteed!
10