PCE Instruments PCE-DFG N TW Serie Bedienungsanleitung Seite 23

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.2
Settings
When you are in measurement mode, press the OK key to enter the settings screen which is
divided into 2 pages:
Page 1
In order to change settings, select the menu item with the arrow keys and confirm with the OK
key. The values can then be changed by means of the arrow keys. Then press "OK" to confirm
the settings or the Back key to discard.
Function
Display Unit
Force Area
Zero Tracking
Sampling Speed
Calibrate Grav
Alarm Upper
Alarm Mode
Description page 1
The display unit can be set to: „Nm", „lbfft" or „kgfm"
The force area can be set to a value between 999.99cm
and is included in the calculation if the display unit selected is „kPA"
(important for the accuracy).
For zero tracking, you can set:
„Off", „0.1 N", „0.2 N", „0.3 N", „0.4 N", „0.5 N"
Values below the value set are automatically excluded before the zero
point stabilises. After stabilisation of the reading, the sampling rate will
be 1 x per second. Deviations from the measured value which are below
the set value are automatically excluded in order to keep the displayed
values.
You can set how many measurements are taken per second. A value
between 6 and 1600 Hz can be set here.
Note: The higher the sampling rate, the lower the accuracy will be.
Higher sampling rates are suitable for dynamic measurements whereas
lower sampling rates are more suitable for static and slow
measurements.
Enter the gravity at the place of calibration.
The upper alarm can be set to +/- 9999.9.
You can choose "Within" (within alarm limit), "Beyond" (outside alarm
© PCE Instruments
Page 2
20
2
2
and 0.01cm