How to Clean and Disinfect
Warning
Risk of personal injury
Caution
When
Steps
Frequency
Do not use cleaning agents containing phenol, as phenol may cause
skin burns if not rinsed thoroughly and does not contain sufficient
cleaning or disinfectant properties.
1. The camera is designed for easy cleaning and disinfection.
2. Before cleaning the camera, disconnect it from the power source.
3. Cleaning procedures should only be carried out by personnel
familiar with camera operation.
4. Do not spray cleaning fluid, especially water, into any internal
electrical equipment or parts of the camera to avoid possible short
circuits, corrosion, malfunction and electrical shock hazards to
users or service personnel.
5. Corrosive cleaning agents may cause discoloration or damage the
camera. Before using any cleaning agent, test it in an
inconspicuous area.
Clean the camera before and after use.
1. Disconnect the camera from a power source.
2. Cleaning personnel must wear cleaning gloves.
3. Before using cleaning alcohol, make sure it has not expired.
4. Use a cleaning cloth dampened with cleaning alcohol (75%).
5. Wipe any surface that may come into contact with the camera.
6. After wiping the camera, avoid contact with bare hands.
No regular cleaning is required. Clean before and after use.