Herunterladen Inhalt Inhalt Diese Seite drucken

Balluff BMD 1H Serie Betriebsanleitung Seite 114

Vorschau ausblenden Andere Handbücher für BMD 1H Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
6 Set up with the display and adjustment module
Setup - Current output
Min./Max.
Setup - False signal sup-
pression
32
In the menu item "Current output Min./Max.", you determine the reac-
tion of the current output during operation.
The default setting is min. current 3.8 mA and max. current 20.5 mA.
High mounting sockets
Vessel internals such as struts
Note:
A false signal suppression detects, marks and saves these false
signals so that they are no longer taken into account for the level and
interface measurement. We generally recommend carrying out a false
signal suppression to achieve the best possible accuracy. This should
be done with the lowest possible level so that all potential interfering
Proceed as follows:
If the probe is covered, enter the actual distance from the sensor to
the product surface.
All interfering signals in this section are detected by the sensor and
stored.
Keep in mind that with covered probe only false signals in the uncov-
ered area of the probe are detected.
Note:
Check the distance to the product surface, because if an incorrect
(too large) value is entered, the existing level will be saved as a false
signal. The level would then no longer be detectable in this area.
If a false signal suppression has already been created in the sensor,
the following menu window appears when selecting "False signal
suppression":

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis