Injector control A
Care and Maintenance
Disconnect the unit from mains supply
before executing care and maintenance!
General notes
Repairs on the Injector control A may only be exe-
cuted by experts. Disconnect the unit from mains
supply at any rate.
Dreve company is only responsible for the safety,
reliability and performance of the unit, if:
• the unit is exclusively used for the corresponding
field of application.
• operation is only carried out by accordingly trained
personnel.
Placing the product out of operation
The DIRECTIVE 2012/19/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 4 July 2012
on waste electrical and electronic equipment (WEEE) lays down measures to protect the environment
and human health. Waste electrical and electronic equipment marked as such may not be disposed
of as unsorted municipal waste within the European Union. For detailed information regarding the
professional disposal of WEEE please contact Dreve or your distributor.
20
Cleaning is effected with a mild cleaning liquid. The
cleaning liquid must not get into the unit.
• the unit is operated in accordance with these
instructions for use.
• extensions, new adjustments, changes or repairs
are executed by persons authorised by the com-
pany.
As we are currently developing our products, we
reserve the right for technical changes.
Trouble shooting
Problem
Cause
Device shows no function
Accu not charged
Device shows no function, accu
Accu defect
loaded
Mains adapter of the charger is
Accu defect
in the socket, accu does not load
Mains adapter of the charger
defect
Cable connection of mains
adapter and Injector control A
is defect
Socket without voltage
Unit produces the material slowly
Accu has been emptied
Motor runs and then stops
Double cartridge has been emptied
Motor runs and then stops
Cured material in old mixing
canula
Unit does not produce, motor
Upper part of the unit has not
runs hearably
snapped into final position,
coupling does not grip
Tensioning band is torn
Unit does not produce, motor
runs hearably
Emptied double cartridge can-
The piston guide is not
not be taken out of the unit
positioned at the rear stud
Aid
Connect Injector control A with
the mains adapter of the char-
ger and charge accu for 2.5 h.
Send the device to the manuf-
acturer
Exchange accu, send the device
to the manufacturer
Exchange mains adapter
Check cable connection
Check power socket
Charge accu
Insert new double cartridge
Apply new mixing canula
Let upper part snap in hearably
The upper part of the unit has
to be demounted by an expert,
the tensioning band has to be
exchanged
Pull piston guide at the opera-
ting lever up to the rear stud
21