AGRETO HFM II
23 Cleaning and maintenance
After making measurements with high humidity, dry the dip until the "L"
symbol appears on the display.
Clean the tip and the lance of the instrument before the season and
when visibly soiled with fine steel wool and / or alcohol.
Store the device on a dry and clean place.
If you are not using the unit for a long time, it is advisable to disconnect
the battery.
24 Troubleshooting
24.1 Battery is empty
If the device displays the battery symbol, the battery needs to be replaced. Any
displayed measurement values can be wrong.
24.2 Wrong measurement values
If you think the displayed measurement values are not correct proceed as
follows:
Note the information in the section "Factors influencing moisture
measurements"
Note the fact that even in dry bales moist areas may occur and / or
locations with higher temperature.
Make sure that the tip and the probe are dry. In the initial state on the
display "L" must appear. If a value is displayed without insertion of the
lance, then the tip is not dry.
Make sure that the tip and the lance are not contaminated.
For temperature measurements, note that the probe must remain in the
material as long as the temperature display does not change.
24.3 Damages on the device
Please contact the manufacturer or your dealer