Important:
If the measured value exceeds the upper display range of the HDA 5500, then the
measured value can no longer be displayed. The upper display range flashes.
If the measured value is below 1 % of the lower display range, then the measured value
can no longer be displayed. The lower display range flashes.

9.2 Special display

If a measured value is set as the primary display, then by pressing the
following values can be displayed separately. These values are each displayed for approxi-
mately 3 s:
Lowest value
Peak value
Switch point(s)
Sensor 1, 2 or 3
The relevant lowest and peak values are reset by pressing the
ously during normal operation. If the reset has been successful, rES (reset) is displayed to
confirm this.

10 Output response

The HDA 5500 has a through analogue output, the signal of which is proportional to an input
signal. In addition, depending on the model and type, the HDA 5500 is available either
without switching outputs, or with two or four switching outputs.

10.1 Analogue output

On the analogue output a 4..20 mA or 0..10 V signal is available. The signal is proportional to
the input signal of one of the sensors or of the differential value. Both the allocation of the
output signal to sensor or differential value and the type of signal can be adjusted during
normal operation.
For setting „diFF" (Analogue output allocated to differential value (sensor 1 – sensor 2))
reference for values of sensor 2 = values of sensor 1 is an analogue output signal of 12 mA
or accordingly 5 V. From there values of sensor 2 < values of sensor 1 are shown with
signals of 0...12 mA or accordingly 0...5 V and values of sensor 2 > values of sensor 1 are
shown with signals of 12...20 mA or accordingly 5...10 V.
10.1.1 Analogue output set to 4..20 mA
Connection example for analogue output set to 4..20 mA:
I
= 4..20 mA
out
Analog. Out
= 400 Ω
RL
max
All manuals and user guides at all-guides.com
AGND
User Manual
or
keys the
and
keys simultane-
- 13 -