2.
Check all castors are free to rotate and that
brakes lock correctly (see section 5.1). If any of
the brakes fail to operate, stop using the chair
immediately.
3.
Check the Pommel locks in place and does not
slide out when pulled (see section 4.4, 5.7).
4.
Check for any damage to the soft moulded parts
(seat liners, headrest, armrests, pommel). If any
of these parts are damaged, they should be
replaced.
7
6
5.
Check the Tilt-in-Space mechanism for smooth
operation. Make sure it locks fully when the red
push buttons are released (see section 5.5).
Check the assembly for hydraulic fluid leaks.
Stop using the chair immediately if the
mechanism becomes difficult to operate, fails to
lock correctly or leaks hydraulic fluid.
6.
Check the security of all adjustment knobs
and fasteners. Pay particular attention to the
armrests (see section 4.7), headrests (see
section 5.4) and anterior supports (see section
6.2).
7.
Check all straps and harnesses for any signs
of damage to the webbing or to the buckles.
Damaged items must be replaced immediately.
Please contact NRS Healthcare Customer Services
to report a fault or book an inspection by a service
technician.
Refer to section 13 for a detailed list of spare parts.
7.3 Annual Inspection
We recommend this product is checked annually by
a qualified technician for safe operation and repair of
any damaged or missing parts.
7.4 Cleaning & Disinfection
The chair should be cleaned at regular intervals,and
dried fully after each use.
Cleaning the chair is an ideal opportunity to identify
any signs of wear or damage (see section 7).
Regular cleaning is also essential to minimise
risk of infection transmission between users and
carers. This is particularly important in a multi-user
environment.
Particular attention must be taken to clean the
commode pan / chair interface, between the liners,
pommel, seat moulding and gas strut.
WARNING: There is a risk of infection
transmission to the carer when cleaning the
chair. Protective equipment should be worn
during any cleaning or maintenance
7
7.5 Lap-strap and Harness Cleaning
7.6 Cleaning the Gas Strut
Note: We recommend weekly cleaning of the gas
strut mechanism.
Note: For hard-water areas we recommend you
clean with filtered / purified water.
1.
2.
10
Clean regularly using a general purpose, neutral
detergent and warm water.
Please check the suitability of the cleaning
product before use. Do not use abrasive cleaners
or cloths as these could permanently damage
the surface and invalidate any warranty.
To disinfect, wipe with a dilute bleach solution,
and rinse thoroughly with fresh water.
Check the chair for functionality after cleaning
the chair.
Clean with sponge and warm soapy water
Do not use bleach or solvents
Do not steam clean, tumble dry or dry clean.
Raise the plastic cover on the gas strut (pulling
upwards) and clean the lower part of the gas
strut with water. Lower the plastic cover and
push down to secure.
Repeat the process for the opposite side.