LB 452
2.4. Electrical Data Bus Module
Elektrische Daten Busmodul
2.4.
Electrical Data, each Measurement Channel
Elektrische Daten, jeder Messkanal
Bus Module
(Option)
Busmodul
(Option)
GSD file is provided. Transmission of the following data:
•
Cyclic output data: Fill level, detector value (CPS)
•
Detector temperature, currently selected calibration curve
•
system status RUN/ERROR/WARNING/STOP)
•
Cyclic input data: Error confirmation, calibration curve, full and
empty adjustment
Profibus DP:
•
Complete Profibus DPV1 Slave according to IEC 61158
•
Automatic Profibus baud rate recognition (9,600 bit/s - 12 Mbit/s)
•
Standardised Profibus RS-485 port
•
Integrated isolation with DC/DC converter and opto-coupler
•
Profibus connection via 9-pole D-Sub socket
Profinet IO:
•
Complete Profinet IO device implementation (Slave) with RT clas-
sification
•
Fast Ethernet transfer 100 MBit/s in full duplex operation
•
Integrated isolation
•
Profinet IO connection via RJ45 standard socket
GSD-Datei wird zur Verfügung gestellt. Übermittlung folgender Da-
ten:
• Ausgabe zyklischer Daten: Füllstand, Detektorwert (CPS), Detektor-
temperatur, aktuell ausgewählte Kalibrierkurve,
und Systemstatus (RUN/ERROR/WARNING/STOP)
• Eingabe zyklischer Daten: Fehlerbestätigung, Kalibrierkurve,
• Voll- und Leerabgleich
Profibus DP:
• Kompletter Profibus-DPV1-Slave nach IEC 61158
• Automatische Profibus-Baudratenerkennung (9.600 bit/s - 12 Mbit/s)
• Standardisierte Profibus-RS-485-Schnittstelle
• Integrierte Potentialtrennung mit DC/DC-Wandler und Optokoppler
• Profibus-Anschluss über 9-polige D-Sub-Buchse
Profinet IO
• Komplette Profinet-IO-Device-Implementierung (Slave) mit RT-Klas-
sifikation
• Fast-Ethernet-Übertragung 100 MBit/s im Full-Duplex-Betrieb
• Integrierte Potentialtrennung
• Profinet-IO-Anschluss über RJ45-Standardbuchse
47344TI Rev.05, 05/2022
Technical Information
Technische Information
Fehlernummer