LB 452
2.3. Electrical Data Extension Module
Elektrische Daten Erweiterungsmodul
2.3.
Electrical Data, each Measurement Channel
Elektrische Daten, jeder Messkanal
Extension Module
(Option)
Erweiterungsmodul
(Option)
4-20 mA current output with option of activating current output signal
monitoring:
potential-free
Switchable source (max. Impedance 500 Ω) or sink (max. DC 24 V,
500 Ω; min. DC 12 V, 250 Ω) mode
Adjustable fault current in the event of an error: 2 mA, 22 mA or
freezing of last value
Can be configured for:
Fill level (additional time constant, freely adjustable)
Detector temperature
Detector count
2 digital inputs:
potential-free, for ext. Choice of up to 4 calibration curves
2 relay outputs for alarm signalling:
Alarm relay 1: Single pole double throw (SPDT)
Alarm relay 2: Single pole single throw, normally open (SPST NO)
Max. of 33 VAC, 46 VDC, 5 A for each, non-inductive
Can be configured for alarm signals:
Max. level
Min. level
Detector temperature
Digital input confirmation
Pulse output:
• Detector pulses looped, max. 12 V amplitude
Fuses:
• 2 x (Si3 and Si4) 5A/T/250 V according to IEC60127-3
• 1 x (Si5) 50 mA/T/250V according to IEC60127-3
4-20 mA Stromausgang mit Option zur Aktivierung der Stromaus-
gangssignalüberwachung:
• potentialfrei
• Umschaltbare Quelle (max. Impedanz 500 Ω) oder Senke (max. DC
24 V, 500 Ω; Mindest. DC 12 V, 250 Ω) Modus
• Einstellbarer Fehlerstrom im Fehlerfall: 2 mA, 22 mA oder Einfrieren
des letzten Wertes
Konfigurierbar für:
• Füllstand (zusätzliche Zeitkonstante, frei einstellbar)
• Detektortemperatur
• Detektoranzahl
47344TI Rev.05, 05/2022
Technical Information
Technische Information
4